Keypad Test

The keypad test ensures that all buttons on the membrane keypad are working. Press Start to
begin the test which requires the user to press each button once, within a 20 second timeframe
(Figure 6-11).
Figure 6-11: Keypad Test
The Test LED button will test just the LED, and ensure that the color of the light on the screen
matches the color of the LED itself.

Odometer Calibration

This test ensures that the odometer is calibrated properly for distance. Even though the
odometer is calibrated in the factory, you must calibrate it periodically to ensure accuracy.
Choose a set distance from the screen, press Start and then move the system that exact
distance, using a measuring tape or other known distance indicator. Press Finish when you
have travelled that distance (Figure 6-12).
Figure 6-12: Odometer Test. Image on the left shows the screen where the user must select a
distance. The image on the right instructs you to move that distance then press Finish.