5.4 Display Unit Problem
While the Display Unit is weatherproof and fairly rugged, it should
be handled in much the same way a notebook computer is. If the
Display Unit does not power up, there may be a problem with the
CPU or the storage media. Always make sure that the access port
on the top of the display unit is closed and sealed.
5.5 Test Line
One of the best ways of detecting problems with the system is,
shortly after receiving the system and getting comfortable with its
operation, collect a line of data at a convenient, easily accessible
location. The line does not have to be too long but one screen (16
meters or 50 feet) is a good guide. This data line should be saved
electronically and perhaps plotted out on paper and dated. The test
line could be collected say, every 6 months and, by reviewing the
previous data, system problems can be detected early. As well, if
there is a suspected problem with the system, this test line could
be collected and compared with earlier tests.
5.6 Contacting the Vendor for Service
When returning the Vendor, have the following information available:
1.
GPR sensor Serial Number displayed at the top of the System
Settings Screen.
.
A brief description of when the error is happening and the
operating conditions (temperature, humidity, sunshine, system
settings, etc.).
6 RD1000™ Operation Manual