Message Circuit Noise Accuracy
SPECIFICATION
TRANSMITTER; Quiet Terminated
RECEIVER
Filters: C-Message, 3 kHz Flat, 15 kHz Flat, Program and 50 kBit
Range
600, 900 and 12000
: 0 to 100 dBm
135 and 1500
: 7 to 100 dBm
Accuracy
: 10 dBm to 100 dBm +1 dBm
: 0 dBm to 10 dBm ±3 dBm
DESCRIPTION
The Message Circuit Noise accuracy is checked using an AC Calibrator and
Decade Transformer. The Decade Transformer attenuates a test signal from
the AC Calibrator.
EQUIPMENT
AC Calibrator
: DATRON 4707A
Decade Transformer
; Model DT72A DEKATRAN
PROCEDURE
CAUTION
Ensure that the AC Calibrator OUTPUT is switched off or set to the minimum level
_ before proceeding with the following procedure.
1
Connect the equipment as shown in the following figure;
AC CALIBRATOR
DBCADE TRANSFORMER
HP 493LA
5-34
Performance Tests