3 Set the HP 4934A as follows;
Measurement
: LEVEL FREQ
Receiver (RCV) Impedance
:
6
OOQ
1
Display Mode
: RCV
4
Check the HP 4934A receiver frequency reading is between 19 Hz and 21 Hz.
5
Set the AC Calibrator frequency to each of the values given in the following table
(as observed on the Frequency Counter) and check the HP 4934A receiver
frequency reading.
AC Calibrator Frequency
HP 4934A Receiver Reading
1,000 Hz
999 to 1001 Hz
10,000 Hz
9,995 to 10,005 Hz
50,000 Hz
49.99 to 50.01 kHz
100,000 Hz
99.99 to 100.01 kHz
110,000 Hz
109.99 to 110.01 kHz
Performance Tests
5-31