Additional instruction for Physician and Technician
1. To Select CPAP or APAP Mode
"MENU"
1.
Press
"UP"
2.
Hold the
for one second. Meanwhile, the LCD screen
allow you to switch the mode from CPAP to APAP.
"UP"
3.
Press
or
4.
After selecting the mode, press
"START/STANDBY"
5.
Press
automatically go back to standby screen 5 seconds later.
2. To Set the Pressure
1. Press
"MENU"
mode.
"UP"
2. Hold the
for one second. Meanwhile, the LCD screen
allow you to adjust the therapy pressure from 4 to 20 cm H
"UP"
3. Press
or
increment or decrement of pressure setting is 0.5 cm H
4. After selecting the prescribed pressure, press
"START/STANDBY"
5. Press
automatically go back to standby screen 5 seconds later.
6. In APAP mode, follow same instructions to set the initial pressure (P
(P
), and minimum pressure (P
max
NOTE–If you try to set an maximum pressure lower than P
automatically to be set lower than P
3. To Re-set the Compliance Meter & Total Meter
"MENU"
1. Press
standby screen.
"UP"
2. Hold the
for one second. Meanwhile,
"CLEAR "
and
"UP"
3. Press
or
compliance meter record will be erased so the value becomes 0.0 hr.
"MENU"
4. Press
5. Press
"START/STANDBY"
automatically go back to standby screen 5 seconds later.
NOTE – When the compliance meter has been cleared, the total meter will also be cleared
at the same time, and vice versa.
XT Auto CPAP System
CPAP Model No.: 9S-005720
(Do not distribute to patients)
[Mode
to select
"DOWN"
and
button, and then simultaneously press the
"DOWN"
button to select CPAP or APAP mode.
"MENU"
button to go back to standby screen or leave the device to
to select
[P
XX.XcmH
"DOWN"
and
button, and then simultaneously press the
"DOWN"
button to increase or decrease the pressure setting. The
button to go back to standby screen or leave the device to
min
[TM XXXX.X hr] or [CM XXXX.X hr]
to select
"DOWN"
and
button, and then simultaneously press the
[TM XXXX.X hr] or [CM XXXX.X hr]
will show on the LCD display.
"DOWN"
. The LCD screen will show
to confirm the reset function.
button to go back to standby screen or leave the device to
CPAP]
menu while in the standby screen.
[Mode
to confirm.
O]
menu while in the standby screen in CPAP
2
[P
XX.XcmH
2
"MENU"
to confirm.
) .
.
max
"CLEAR OK"
"MENU"
CPAP]
should start blinking to
"MENU"
O]
should start blinking to
2
O.
2
O.
), maximum pressure
init
and P
, P
init
min
init
menu while in the
"MENU"
will start blinking
and the total meter or
button
button
and P
will
min
button
Need help?