iCH Auto/ English
NOTE: Make sure the silicon connector on the inlet connector fits securely over the CPAP
device air outlet.
CAUTION: Do not overfill the water chamber. Damage to the device may occur.
CAUTION: Use only room temperature water; do not fill the chamber with hot or chilled water.
CAUTION: Do not turn the heater on without the water chamber installed
CAUTION: When installing the water chamber, do not allow water to spill into the device.
CAUTION: Do not move the device when the chamber has water in it.
3.3 Setting Up
1. Place the device on a flat surface and at a lower level than your sleeping position.
2. Connect one end of the air tubing firmly to the air outlet on the back of the device, as shown in
Figure 3-8.
3. Connect the other end of the air tubing to the mask system. Put on the mask and headgear
according to the mask instruction manual.
4. Plug the socket end of the AC power cord into the power supply adapter. Plug the pronged end of
the AC power cord into a main electrical outlet.
5. Plug the power supply cord's adapter connector into the power inlet on the right side of the
device, as shown in Figure 3-9.
6. Once the power supply cord's adapter connector is plugged into the power inlet, the CPAP
system is in ready to operate position ("STANDBY" sign appears in LCD display)
NOTE: The AC power cord also serves to disconnect the device.
NOTE: Do not position the equipment so that it is difficult to operate the disconnecting
device.
Figure 3-8
Figure 3-9
7