To correct these types of errors we can modify the pulses per litre factor programmed in the
instrument.
Example - Shortage of volume
If we have a converter which specifies a k factor = 1.985, and when we check the volume of
a batch, we find that instead of having 100 litres as programmed, we have 105 litres (5%
more), we must apply the following correction:
F
= k factor (original pulses per litre)
k
V
= Expected Volume
Vr
= Real Volume
F
n = New k factor
k
6.2

Change of units of measurement

In some cases we need to change the measurement units, for example, instead of working
in litres we need to work in kilograms. In this case we will need to know the density of the
liquid (ρ).
To change from litres to kg we must divide the pulse per litre factor by the density of the
liquid to obtain the new factor for programming the CIP II. For example, if the liquid has a
density of 0.9 kg/litre, the meter gives us 200 pulses per litre and we must count in kg, we
will program the CIP II using 222.222 as the new pulses per litre factor to be able to count
directly in kg.
F
= Pulses per litre factor to count in kg
k
F
= Original pulses per litre factor
ρ
= Density of the liquid in kg/litre
7

TECHNICAL CHARACTERISTICS

Power supply
CR2450 lithium battery
Nominal voltage:
Capacity:
Power consumption:
Totalizer counter
N. of digits:
Size of the digit:
Reset:
Partial counter
N. of digits:
Size of the digit:
Reset:
General characteristics
Ingress protection:
Ambient temperature range:
Battery life:
= 1.985
= 100
= 105
= ? (1.887)
3 V
560 mAh
8 μA
7
8 mm
Cannot be reset
5
6 mm
By means of push button or magnet
IP65
0ºC ... +60ºC
More than 5 years