Instructions for Use
www.airliquidehealthcare.com.au

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Presence PR and is the answer not in the manual?

Questions and answers

Summary of Contents for Air Liquide Presence PR

  • Page 1 Presence PR, PRC & PRE Instructions for Use www.airliquidehealthcare.com.au...
  • Page 2: Table Of Contents

    2.1 Connecting to the barbed outlet 2.2 Connecting to the bleeved outlet Troubleshooting guide The Integrated Regulator-Valve is manufactured by Air Liquide Medical Systems, 6 rue Georges Besse 92182 Antony France and is supplied as part of the Presence™ range of products in Australia by: Air Liquide Healthcare Pty.
  • Page 3 Presence™ PR, PRC and PRE are ready to use medical oxygen cylinders complete with an integrated regulator-valve and flow meter which are intended to supply regulated medical oxygen to a patient or another medical device. They deliver medical oxygen either through a flow regulated barbed outlet or a fixed flow sleeved outlet.
  • Page 4: Inspection Checks To Be Performed By The User Before Initial Use

    (refer Fig 3). If any deviation is observed do not use the cylinder and contact your nearest Air Liquide Healthcare Figure 3: Operating instructions and product office on 1300 360 202.
  • Page 5: Operational Information

    2. Operational information 2.1 Connecting to the barbed outlet Remove the cap and connect the oxygen tubing to the barbed outlet (refer Fig 4). Turn the ON/OFF valve (outer ring) anti clockwise to show the green “I” in the portal on top of the valve (refer Fig 5).
  • Page 6: Connecting To The Bleeved Outlet

    Figure 9: Connection to the PRE Sleeved Outlet Ensure that the extension piece is removed from using extension piece the cylinder prior to returning the empty cylinder to Air Liquide Healthcare.
  • Page 7: Troubleshooting Guide

    Obtain another unit for use. If the unit is required immediately, then twist the shroud back in-line and use as normal. At the completion of use, contact customer service and quarantine the unit for investigation by an Air Liquide Healthcare representative. Damaged sleeved...
  • Page 8 Fax: 08 6389 1199 *All eFax area codes begin with (02). www.airliquide.com World leader in gases, technologies and services for Industry and Health, Air Liquide is present in 80 countries with approximately 65,000 employees and serves more than 3 million customers and patients.