VACUUM TECHNOLOGY OPERATING AND MAINT ENANCE INSTRUCTIONS (Translation of the original instructions) VACUUM PUMPS WITH RECIRCULATING LUBRICATION SYSTEM LB.5WR LC.25WR LB.40WR LB.60WR LC.105WR www.dvp.it info@dvp.it...
Page 2
Please visit our website (www.dvp.it) for further information on the other D.V.P. Vacuum Technology products and to get in touch with our staff.
MAINTENANCE TABLE..............................13 6.2.1 CHECK THE OIL LEVEL AND THE ABSENCE OF WATER VAPOUR CONDENSED IN THE TANK ....14 6.2.2 CHECK THE STATUS OF THE INTAKE FILTER (ONLY LB.5WR) ............... 14 6.2.3 CHANGE OIL................................14 6.2.4 INTAKE FILTER CARTRIDGE REPLACEMENT (ONLY LB.5WR) ................ 14 6.2.5...
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR INTRODUCTION GENERAL INFORMATION This manual is meant to provide you with important information for the safety of persons involved in pump use and maintenance. This manual, originally written in ITALIAN, is an integral part of the pump and must be preserved with care for the life of the pump itself.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR INFORMATION PLATE All pumps are equipped with an identification plate that contains the manufacturer's name, address, CE marking and technical data of the pump itself. VACUUM TECHNOLOGY WARNING Removing or tampering with the identification plate is strictly prohibited.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR Electrical hazard Electrical equipment in the pump includes live parts which, upon contact, can cause serious damage to persons and property. Any kind of intervention on the electrical system must be performed by qualified personnel.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR 3.3.3 Model: LB.40WR – LB.60WR Only on "FKM" versions Intake outlet Terminal board Information Attachment Attachment plate point point viewer Motor Motor fan Oil sight Oil filler plug Oil drainage...
STORAGE Drain the oil inside the pump and close the intake and outlet with the appropriate protections and, only for LB.5WR, disassemble the filter at the intake using the relevant connector (see paragraph 3.3.1). The pumps are to be stored in their packaging and stored in covered, dry, protected places that are not exposed to bright sunlight, with temperatures in the range indicated in the table of technical specifications.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR Lighting All areas must be illuminated evenly and sufficiently to ensure all operations included in this manual and must be without shadows, reflections, glare or eyestrain. PUMP INSTALLATION To ensure perfect pump operation, house and place it according to the following conditions: •...
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR WARNING Always install an intake filter, especially if the pump is to work with air that might contain foreign matters. Pump outlet gases must be treated in such a way that they do not contaminate the working environment and the surrounding atmosphere.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR OPERATING INSTRUCTIONS OPERATION Checks to be performed before start-up: • The pump is supplied without any oil in it. Use the supplied oil or an oil lubricant indicated on the identification plate of this pump (see ch.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR 5.1.5 STOP The pump must be stopped by cutting off the power supply. WARNING If the pump is to be powered off, let it run with closed intake, isolated from the user system, for about 30 minutes first.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR 6.2.1 CHECK THE OIL LEVEL AND THE ABSENCE OF WATER VAPOUR CONDENSED IN THE TANK Check that oil level is at mid-range of the oil level indicator light or the oil viewer. If not, see instructions in paragraph 5.1.1.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR SPARE PARTS Use Original Spare Parts to replace pump parts. When purchasing spare parts, always quote the serial number and model of the pump (these can be found on the identification plate) as well as the spare part purchase number.
Operating and Maintenance Instructions LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR TROUBLESHOOTING DAMAGE CAUSE REMEDY No voltage Provide power supply Thermal switch has tripped Identify reason and activate switch Room temperature is too low Restore room temperature to allowed range...
Page 19
DVP Vacuum Technology s.r.l. will not be held responsible for any direct or indirect losses caused by its failure to respect the assistance times and will not have any responsibility or contractual or civil obligation for product faults or for failure to repair the faults in a reasonable period of time.