en - Areas of application
A 805 holder for
stapler S/Si
46
The rinse connection on the shaft of the stapler must face upwards.
Press in the middle motor connection on the upper end of the hous-
ing slightly and turn the shaft until the rinse connection is facing up-
wards.
Connect the hose adapter to the rinse connection on the shaft of
the stapler.
The rinse pipe must click into place in the rinse connection. To do
this:
If required, turn the clamp around the shaft until the rinse pipe
clicks into position.
The clamp is used to secure the hose adapter.
If the hose adapter has come loose from the rinse connection on
the stapler during reprocessing, it must be reprocessed again.