深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Safety Precautions The following are general safety precautions that are not necessarily related to any specific part or procedure, and do not necessarily appear elsewhere in this publica- tion. These precautions must be thoroughly understood and apply to all phases of operation and maintenance.
This symbol appears on the equipment and indicates the requirement for separate collection of discarded electrical and electronic equipment in accordance with the European Union Directive 2002/96/EC. Refer to the Bird web site (See “The Bird Web Site” on page 22) for more information. Warning Statements The following safety warnings appear in the text where there is danger to operat- ing and maintenance personnel, and are repeated here for emphasis.
Page 4
Maximum power is 1,250 W when the blower is not running. If the indicator light should turn off, immediately reduce RF power to less than 1,250 W. On page 13. CAUTION Use only Bird coolant, P/N 5-1070, to prevent damage to the load. On page 17. 深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com...
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Safety Statements USAGE ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. EL USO DE ESTE INSTRUMENTO DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
Page 6
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 SERVICE SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE - TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED TO DO SO. SERVICIO LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO.
Page 7
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 CONNECT INTERLOCK TO TRANSMITTER BEFORE OPERATING. BRANCHER LE VERROUILLAGE À L'ÉMETTEUR AVANT EMPLOI. CONECTE EL INTERBLOQUEO AL TRANSMISOR ANTES DE LA OPERACION. VOR INBETRIEBNAHME VERRIEGELUNG AM SENDER ANSCHLIESSEN. PRIMA DI METTERE IN FUNZIONE L'APPARECCHIO, COLLEGARE IL DISPOSITIVO DI BLOCCO AL TRASMETTITORE.
The remainder of this instruction book is divided into Chap- ters and Sections. Chapter Layout Introduction - Describes the features of the Bird SignalHawk, lists equipment supplied and optional equipment, and provides power-up instructions. 深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com...
Page 9
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Theory of Operation - Describes how to use the SignalHawk’s PC Tool for the Vec- tor Network Analysis function and how to transfer readings from the SignalHawk to the computer and back again. Operation without Blower - Describes procedures require for operating the load resistor.
Chapter 1 Introduction Bird 889x loads are portable, 50 ohm, coaxial RF transmission line termina- tions, designed for frequency ranges of DC – 2.4 GHz. 889xD loads are designed for use in the UHF (470 – 860 MHz) band but are otherwise identi- cal.
Theory of Operation Load Resistor Bird 889x loads consist of a thin-film-on-ceramic resistor immersed in a dielec- tric coolant. The resistor, individually selected for its accuracy, is enclosed in a special housing. When surrounded by the coolant, this produces a uniform, practically reflectionless line termination over the specified frequencies.
Page 15
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Theory of Operation 深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356...
This load is designed for operation in a horizontal position only, with the vent plug up. Do not use in any other manner. Bird 889x Loads are equipped for either portable use or fixed installation. The mounting brackets on the front and rear faces have four mounting slots arranged in a 4 ⁄...
Installation Installing the Thermoswitch Bird 889x series Loads can be equipped with an optional interlock thermo- switch, P/N 8890-008. It is normally closed, opening at 236 °C (457 °F), with a rating of 10A @ 120Vac and 5A @ 230Vac.
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Installation 6. Securely solder the control switch leads to the lugs (G) of the connector base. Note: The ring-nut (A) must be in place over the base plug (D) with the knurled end facing out. 7.
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Installation Using a Swivel Flanged Coupling To couple the swivel flange with a flanged RF transmission line, use an appro- priate coupling kit. Refer to Figure 5 while following the instructions. Figure 5 Swivel Flanged Coupling RF COAXIAL LINE BOLT...
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Installation Using an AC Power Hookup WARNING Turn off AC power and RF power when attaching the power cable. CAUTION Check the local electrical code for proper AC hookup prior to operation of the unit.
Chapter 4 Operating Instructions, No Blower Normal Operation Bird 889x–300 loads have no indicators or operating controls. They require no special operating procedures or surveillance when their performance limits are not exceeded. Follow the instructions for the specific transmitter equipment.
RF power is being applied. Leaking RF energy is a potential health hazard. Blower Controls Bird 889x–310 and –320 Loads are equipped with a control switch and indica- tor light on the front of the blower assembly. The switch is labelled MANUAL/ AUTOMATIC.
Chapter 6 Maintenance This chapter covers cleaning, inspection, trouble-shooting, and specifications for the Bird 889x Loads. WARNING Never attempt to connect or disconnect RF equipment from the transmission line while RF power is being applied. Leaking RF energy is a potential health hazard.
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Maintenance Maintenance Cleaning Cleaning the Load Resistor The outside surface of the instrument should be wiped free of dust and dirt when necessary. Excessive dust on the cooling fins will interfere with heat dissipation. Clean the RF connector, both metallic and insulating surfaces, with a dry, non-residue forming solvent.
1. Remove the load resistor (Refer to “Load Resistor” on page 18). CAUTION Use only Bird coolant, P/N 5-1070, to prevent damage to the load. 2. Inspect the coolant. The coolant should be clear, with a faint yellow tinge, and have a slightly sweet smell.
The potential for electric shock exists. Replacing the RF Connector ONLY the Model 8890 has a special Bird “QC” connector which allows easy changing of the RF connector. This does not disturb the coolant seal or affect the electrical continuity of the load.
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Maintenance Replacing the Load Resistor 1. Remove the vent plug and install the shipping plug. 2. Stand the unit on its back with the connector end up. Note: In this position there is no danger of the coolant pouring out through the socket plug hole.
LIGHT Storage and Shipment Storing the Load Resistor Cover Bird 8890 Loads before storing to keep out dust and dirt. It is not necessary to install the shipping plug. Store in a dry, dust-free environment where the ambient temperature will remain between –40 and +45 °C (–40 to +113 °F).
If you need to return the unit for any reason, request an RMA through the Bird Technologies website (link shown below). All instruments returned must be shipped prepaid and to the attention of the RMA number.
Page 33
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Maintenance Specifications Frequency Range 8891D300, 8892D300, 8892D13-30, 8892D320 470 – 860 MHz All other models DC – 2.4 GHz Power Rating 2500 W continuous duty † Peak Power for Pulse Width 1 µs 150 kW 10 µs 115 kW...
深圳市新朗普电子科技有限公司 TEL:0755-83986300 FAX:0755-88850515 www.xLp17.com Email:fuhao@17Lp.com QQ:55178055 285480356 Limited Warranty All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one year, unless otherwise specified, from date of shipment and to conform to applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be other than as warranted;...