Power Requirements
Power Rail
+3.3 V
+5 V
1
+12 V
+5 V
AUX
1
Not including the power consumed by an active cable. NI copper cables (785550-0x) are
passive.
Environment
Maximum altitude
Pollution Degree
Indoor use only.
Operating Environment
Ambient temperature range
Relative humidity range
Storage Environment
Ambient temperature range
Relative humidity range
Caution
device is completely dry and free from contaminants before returning it to service.
30 | ni.com | MXI-Express Gen-3 x16 User Manual
Typical Current
Clean the PXIe-8399 with a soft nonmetallic brush. Make sure that the
750 mA
100 mA
1100 mA
200 mA
2,000 m
2
0 °C to 55 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)
10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
-40 °C to 71 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
5% to 95%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Maximum Current
1.1 A
200 mA
2500 mA
500 mA