Daily Cleaning Instructions: KL1 
 
Clean the following components at least once daily.  Clean more frequently if your pharmacy 
dispenses many dust‐generating medications or if you experience counting errors. 
 
Materials:  70% isopropyl alcohol or mild soapy water (e.g., dish soap), lint‐free cloth or 
paper towels. Do not use facial tissue, cotton balls or cotton swabs. 
 
Before starting, turn off the unit and unplug it. 
 
1. Funnel 
A. Remove the funnel from the unit.  
B. Clean outside and inside thoroughly with alcohol or mild soapy water. 
 
 
2. Glass Detector Windows 
The unit's counting sensors are located behind the removable inside channel assembly. 
Most counting errors are caused by pill dust accumulating on the edges and corners of 
the windows on the channel assembly. Clean the windows well. 
A. Unscrew the two round thumb screws on top of the unit. Remove the  
square guard. 
B. Lift out the entire inside channel assembly. Do not remove the individual glass panes 
attached to the inside channel assembly. 
C. Wipe off excess pill dust on the inside of the four clear detector windows with  
a dry cloth. 
D. Wet a clean cloth with alcohol or mild soapy water and clean the inside of the four 
glass windows thoroughly to remove all remaining dust and residue. Wipe back‐and‐
forth, not up‐and‐down. Do not press hard as this may break the glass windows. Do 
not remove the individual glass panes. 
E. Fold the cloth to form an edge to clean any edge dust can accumulate, especially corners. 
F. Wipe the glass windows dry with a clean cloth. Hold the windows up to a light to visually 
inspect the glass for any dust, fingerprints, or streaks. Re‐clean if necessary. Wipe dry. 
G. Look down inside the KL1 device and visually inspect the four inside glass windows  
for dust or streaks. If necessary, clean with water and wipe dry. 
H. Replace the inside channel assembly and square guard. Tighten the round thumb  
screws until snug. Do not over‐tighten the thumb screws. 
 
Step A – Remove 
Square Guard 
Steps C‐F, Inside 
Channel Assembly: 
Clean inside and 
outside of 4 glass 
windows without 
removing or 
damaging the panes 
  5DIM‐3802‐DCI   4/2016 

Advertisement

loading
Need help?

Need help?

Do you have a question about the KL1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kirby Lester KL1

  • Page 1 Step A – Remove  Square Guard  2. Glass Detector Windows  The unit’s counting sensors are located behind the removable inside channel assembly.  Most counting errors are caused by pill dust accumulating on the edges and corners of  the windows on the channel assembly. Clean the windows well.    A. Unscrew the two round thumb screws on top of the unit. Remove the   square guard.  B. Lift out the entire inside channel assembly. Do not remove the individual glass panes  attached to the inside channel assembly.  C. Wipe off excess pill dust on the inside of the four clear detector windows with   a dry cloth.  D. Wet a clean cloth with alcohol or mild soapy water and clean the inside of the four  glass windows thoroughly to remove all remaining dust and residue. Wipe back‐and‐ forth, not up‐and‐down. Do not press hard as this may break the glass windows. Do  not remove the individual glass panes.  E. Fold the cloth to form an edge to clean any edge dust can accumulate, especially corners.  Steps C‐F, Inside  F. Wipe the glass windows dry with a clean cloth. Hold the windows up to a light to visually  Channel Assembly:  Clean inside and  inspect the glass for any dust, fingerprints, or streaks. Re‐clean if necessary. Wipe dry.  outside of 4 glass  G. Look down inside the KL1 device and visually inspect the four inside glass windows   windows without  for dust or streaks. If necessary, clean with water and wipe dry.  removing or  damaging the panes  H. Replace the inside channel assembly and square guard. Tighten the round thumb   screws until snug. Do not over‐tighten the thumb screws.    5DIM‐3802‐DCI   4/2016 ...
  • Page 2   If desired, the glass window holder can be removed from the inside channel  assembly for a more thorough cleaning and visual inspection. Gently push down on  the square sliding tube, which spreads the two black clips holding the glass window  holder. Take special care not to break the black clips.    3. Counting Tray   Clean thoroughly with mild soapy water. Do not use alcohol.  Step I – Visually  Inspect Windows      4. Tray Sensor   Located on the KL1 behind the counting tray when the tray is fully inserted. This sensor  tells the unit that the tray is properly seated. With the tray removed, clean the sensor  window with water. Wipe dry to remove streaks.        Tray Sensor    Note about the use of mild soapy water versus isopropyl alcohol: Mild soapy water and alcohol both  provide appropriate sanitary cleansing. Water leaves fewer streaks than alcohol on a Kirby Lester  device's inside glass sensor windows. If streaks are found after cleaning, use just water and wipe dry.  In general, prolonged alcohol exposure causes faster wear‐and‐tear on plastic.     5. Cleaning and Troubleshooting     Problem or Error Code  Probable Cause Possible Solution(s)  E‐2, E‐4, E‐9, O‐S or unit counts by  Tablet dust or residue on glass detector  Clean glass detector windows thoroughly. Or  itself  windows, or replace cracked glass detector  replace cracked glass detector window.  window.  Overcounts  Tablet fragments, or tablet dust on glass ...