Kirby Lester KL15 Series Daily Cleaning Instructions

Advertisement

Daily Cleaning Instructions: KL15 Series 
 
Clean the following components at least once daily.  Clean more frequently if your pharmacy 
dispenses many dust‐generating medications or if you experience counting errors. 
 
Materials: Distilled water, 70% isopropyl alcohol, mild soap (e.g., dish soap), lint‐free cloth or paper 
towels. Do not use facial tissue, cotton balls or cotton swabs. 
 
Before starting, turn off the unit and unplug it. 
 
1. Top Funnel 
A. Remove the top funnel and channel from the unit.  
B. Clean outside and inside thoroughly with mild soapy water or alcohol. 
 
 
2. Glass Detector Windows 
The unit's counting sensors are located behind four glass windows inside the  
square funnel hole. Most counting errors are caused by pill dust accumulating  
on the edges and corners of the glass windows. Clean the windows well. 
A. Wipe off excess pill dust on the four glass windows with a dry  cloth.  
B. Wet a clean cloth with mild soapy water or alcohol and clean the inside of 
the four glass windows thoroughly to remove all remaining dust and residue.  
Carefully wipe back‐and‐forth, not up‐and‐down. Do not press hard as this  
may break the glass windows.  
C. Wipe glass windows dry with clean cloth to remove streaks. 
 
NOTE: If your unit has a clear detector window assembly (not black), clean with eyeglasses cleaner, 
not alcohol. Contact Kirby Lester service for questions at 800.243.2465 or service@kirbylester.com. 
 
3. Counting Tray (KL15e, KL15es, KL15i only) 
A. Carefully remove the clear plastic tray cover. Clean all parts thoroughly with a water‐
dampened cloth (not alcohol). Wipe dry. Replace the tray cover.   
B. Turn the tray over, and wipe the silver sensor strip with a water‐dampened cloth 
(not alcohol). 
 
4. Tray Sensor (KL15e, KL15es, KL15i only)  
A. (Located on the bottom of the unit in back of the counting tray.) This sensor 
detects the tray's silver sensor strip, telling the unit that the tray is properly seated.  
B. With the counting tray removed, tip the unit carefully to access the tray sensor.  
Clean with a water‐dampened cloth (not alcohol).   
 
 
D5AO‐0000‐DCI   3/2015 
2‐piece 
Funnel 
Four Detector 
Windows 
Tray Sensor 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KL15 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kirby Lester KL15 Series

  • Page 1 Daily Cleaning Instructions: KL15 Series    Clean the following components at least once daily.  Clean more frequently if your pharmacy  dispenses many dust‐generating medications or if you experience counting errors.    Materials: Distilled water, 70% isopropyl alcohol, mild soap (e.g., dish soap), lint‐free cloth or paper  towels. Do not use facial tissue, cotton balls or cotton swabs.    Before starting, turn off the unit and unplug it.    1. Top Funnel  A. Remove the top funnel and channel from the unit.   2‐piece  B. Clean outside and inside thoroughly with mild soapy water or alcohol.  Funnel      2. Glass Detector Windows  The unit’s counting sensors are located behind four glass windows inside the   Four Detector  square funnel hole. Most counting errors are caused by pill dust accumulating   Windows  on the edges and corners of the glass windows. Clean the windows well.    A. Wipe off excess pill dust on the four glass windows with a dry  cloth.   B. Wet a clean cloth with mild soapy water or alcohol and clean the inside of  the four glass windows thoroughly to remove all remaining dust and residue.   Carefully wipe back‐and‐forth, not up‐and‐down. Do not press hard as this   may break the glass windows.   C. Wipe glass windows dry with clean cloth to remove streaks.    NOTE: If your unit has a clear detector window assembly (not black), clean with eyeglasses cleaner,  not alcohol. Contact Kirby Lester service for questions at 800.243.2465 or service@kirbylester.com.    3. Counting Tray (KL15e, KL15es, KL15i only)  A.
  • Page 2   5. Black Lower Funnel (KL15df only)   A. Unscrew the black lower funnel. Clean the interior thoroughly with mild soapy water or  alcohol. Dry black funnel’s interior thoroughly.   B. Clean the silver exit gate with mild soapy water or alcohol. Dry thoroughly.  C. Reassemble only when parts are dry.    Note about the use of mild soapy water versus isopropyl alcohol: Mild soapy water and alcohol both  provide appropriate sanitary cleansing. Water leaves fewer streaks on a Kirby Lester device's inside  glass sensor windows than alcohol. If streaks are found after cleaning, use just water and wipe dry.   In general, prolonged alcohol exposure causes faster wear‐and‐tear on plastic.    6. Cleaning and Troubleshooting     Problem or Error Code  Probable Cause Possible Solution(s)  E‐2, E‐4, E‐5, E‐6, E‐9, O‐S  Tablet dust or residue on glass detector  Clean glass detector windows thoroughly.  or unit counts by itself  windows, or possible cracked glass detector  Or replace cracked glass detector window  window.  assembly (incl. 4 windows).  Overcounts  Tablet fragments, or tablet dust on glass  Check counted medications for tablet  detector windows.  fragments. Clean glass detector windows  thoroughly.  Undercounts  Pouring too fast, or pouring tablets directly  Slow pouring speed so the “Pour Slower”  into the funnel hole.  indicator does not activate.  KL15e, KL15es, KL15i  Tray not fully inserted. Tray sensor window ...

Table of Contents

Save PDF