Daily Cleaning Instructions: KL16 Series 
 
Clean the following components at least once daily.  Clean more frequently if your pharmacy 
dispenses many dust‐generating medications or if you experience counting errors. 
Materials: 70% isopropyl alcohol, lint‐free cloth and water. 
Do not use facial tissue, cotton balls or cotton swabs to clean the unit. 
 
Before starting, unplug the unit. 
 
1. Top Funnel 
o Remove the top funnel and middle channel from the unit.  
o Clean outside and inside thoroughly with alcohol. 
 
 
2. Glass Detector Windows 
The unit's counting sensors are located behind four glass 
windows inside the square funnel hole. Most counting errors are caused by pill dust 
accumulating on the edges and corners of the windows. Clean the windows well. 
o Wipe off excess pill dust on the detector windows with a dry lint‐free 
cloth. 
o Wet a clean cloth with alcohol and clean the inside of the four glass 
windows thoroughly to remove all remaining dust and residue. Carefully 
wipe back‐and‐forth, not up‐and‐down.  
o Do not press hard as this may break the glass windows. Broken windows 
are not covered under the warranty. 
o Wipe glass windows dry with clean cloth to remove alcohol streaks. 
 
 
3. Counting Tray (KL16 only) 
o Carefully remove the clear plastic tray cover. Clean all parts thoroughly with a water‐
dampened cloth (not alcohol). Wipe dry. Replace the tray cover.   
o Turn the tray over, and wipe the silver sensor strip with a water‐dampened cloth (not 
alcohol). 
 
 
4. Tray Sensor (KL16 only)  
o (Located on the bottom of the unit in back of the counting tray.) This sensor 
detects the tray's silver sensor strip, telling the unit that the tray is properly 
seated. 
o With the counting tray removed, tip the unit carefully to access the tray 
sensor.  Clean with a water‐dampened cloth (not alcohol).   
 
 
2‐piece 
Funnel 
4 Detector 
Windows 
Tray Sensor
12/2011

Advertisement

loading
Need help?

Need help?

Do you have a question about the KL16 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kirby Lester KL16 Series

  • Page 1 Daily Cleaning Instructions: KL16 Series    Clean the following components at least once daily.  Clean more frequently if your pharmacy  dispenses many dust‐generating medications or if you experience counting errors.  Materials: 70% isopropyl alcohol, lint‐free cloth and water.  Do not use facial tissue, cotton balls or cotton swabs to clean the unit.    Before starting, unplug the unit.    1. Top Funnel  o Remove the top funnel and middle channel from the unit.   2‐piece  o Clean outside and inside thoroughly with alcohol.  Funnel      2. Glass Detector Windows  The unit’s counting sensors are located behind four glass  4 Detector  windows inside the square funnel hole. Most counting errors are caused by pill dust  Windows  accumulating on the edges and corners of the windows. Clean the windows well.    o Wipe off excess pill dust on the detector windows with a dry lint‐free  cloth.  o Wet a clean cloth with alcohol and clean the inside of the four glass  windows thoroughly to remove all remaining dust and residue. Carefully  wipe back‐and‐forth, not up‐and‐down.   o Do not press hard as this may break the glass windows. Broken windows  are not covered under the warranty.  o Wipe glass windows dry with clean cloth to remove alcohol streaks.      3. Counting Tray (KL16 only)  o Carefully remove the clear plastic tray cover. Clean all parts thoroughly with a water‐...
  • Page 2   5. Black Lower Funnel (KL16df only)   o Unscrew the black lower funnel. Clean the interior thoroughly with alcohol. Dry black  funnel’s interior thoroughly.   o Clean the silver exit gate with alcohol and a lint‐free cloth. Dry thoroughly.  o Reassemble only when parts are dry.    6. Cleaning and Troubleshooting     Problem or Error Code  Probable Cause Possible Solution(s)  E‐2, E‐4, E‐5 or E‐6  Tablet dust or residue on detector windows. Clean detector windows thoroughly. E‐9, O‐S, or unit counts by  Tablet dust or residue on detector windows,  Clean detector windows thoroughly. If  itself  cracked detector window, or possible  error persists, contact Service.  detector board failure  Miscounts  Tablet fragments, or tablet dust or residue  Check counted medications for tablet  on detector windows.  fragment(s); clean detector windows  thoroughly.  KL16 Only: Display reads  Tray not fully inserted. Tray sensor window  Fully insert counting tray; clean tray  “‐‐‐“ instead of “000”  is dirty. Silver sensor strip on bottom of tray  sensor window (damp cloth, not alcohol);  with tray pushed in  is peeling or worn.  replace silver sensor strip. If error ...