Daily Cleaning Instructions: KL20
Clean the following components at least once daily. Clean more frequently if your pharmacy
dispenses many dust‐generating medications or if you experience counting errors. Follow the
instructional video directions shown on the screen and in this guide to clean the KL20.Your warranty
does not cover damage due to misuse or abuse.
Materials: Distilled water or 70% isopropyl alcohol, lint‐free cloth or paper towels, and
computer screen cleaner. Do not use facial tissue, cotton balls or cotton swabs.
Before starting, turn off and unplug the unit.
1. Top Funnel
A. Remove the top funnel and channel from the unit.
B. Clean outside and inside thoroughly with mild soapy water or alcohol.
2. Glass Detector Windows
The unit's counting sensors are located behind four glass windows inside the
square funnel hole. Most counting errors are caused by pill dust accumulating
on the edges and corners of the windows. Clean the windows well.
A. Wipe off excess pill dust on the glass windows with a dry lint‐free cloth.
B. Wet a clean cloth with water or alcohol and clean the inside of the four glass
windows thoroughly to remove all remaining dust and residue. Carefully wipe
back‐and‐forth, not up‐and‐down. Do not press hard as this may break the glass
windows.
C. Wipe glass windows dry with clean cloth to remove streaks.
3. Counting Tray
A. Carefully remove the clear plastic tray cover. Clean all parts thoroughly with a water‐
dampened cloth (not alcohol). Wipe dry. Replace the tray cover.
B. Turn the tray over, and wipe the silver sensor strip with a water‐dampened cloth
(not alcohol).
4. Tray Sensor
A. (Located on the bottom of the unit in back of the counting tray.) This sensor detects
the tray's silver strip, telling the unit that the tray is properly seated.
B. With the counting tray removed, tip the unit carefully to access the tray sensor.
Clean with a water‐dampened cloth (not alcohol).
5. Customize Onscreen Cleaning Reminders
You can customize the automatic cleaning reminder function of your unit. See the KL20
operating manual for full instructions on making these adjustments:
Turn the cleaning checklist off
Change the frequency of the cleaning reminder (default 150 scripts)
Add medications that automatically trigger a cleaning reminder (e.g., sulfa)
2‐piece
Funnel
4 Detector
Windows
Tray Sensor