If your control solution test results continue to fall outside the range printed on
the test strip vial, the GlucoDr Plus Blood Glucose Monitoring System may not
be working properly.
- If this is the case, do not use the system to test your blood glucose levels.
- Contact your local distributor.
Record the Control Solution Test Results in the Meter
To record the control solution test results, press any of UP/DOWN button
of the meter to select ATTENTION ICON. Press and release the power button
if selection is completed. The control solution test results will be saved.
Record all the control solution test results with
ATTENTION ICON '
actual blood glucose tests in the meter memory. Recorded
control solution test results will not be calculated in your
result averages.
' to distinguish them from those of