ULTRASONIC PORTABLE MILK ANALYZER
CHAPTER SIX
Error
message
problem /cause
Insufficient
quantity of the
Empty
milk
Sensor
sucked in the
system or air in
the sample
Sucked
Sample
overheated
Overheated
sample
Troubleshuting
Possible
-
- Check if after starting measurement,
sample
- Check if the end of the suction pipe is
- Check the hose connections.
- The sample temperature exceeds the
w w w . m i l k o t e s t e r . c o m
Repair/remedy
Check that there are no air bubbles
in
the sample.
milk sample in the sample holder
decreases. In other case – there is
damage in the suction system.
not above the surface of the liquid
(not dipped enough).
maximum permissible sample's
temperature.
- 23 -