Calibration - Milkotester MASTER MINI User Information

Milk analyzing device
Table of Contents

Advertisement

ULTRASONIC PORTABLE MILK ANALYZER
Press the ↑↓ (up and down) arrows to correct the value with 0.01.
Press ENTER to validate the correction or EXIT to keep the previous
value.

4.3.2 Calibration

MENU > Mode > Calibration.
In order to perform calibration of the device for a certain type of milk /for
example – cow milk/. You need two samples of milk with known values
of their parameters – one with comparatively high percentage of Fat and
one with comparatively low percentage of Fat.
After pressing ENTER on the display appears:
Use the ↑↓ (up and down) arrows and press ENTER to select the type of
milk.
Fat
SNF
Density
Protein
Lactose
Solids
Water
Correction
Of Fat
0.01
Cow milk
Sheep milk
UHT milk
Calibr. With PC
- 14 -
w w w . m i l k o t e s t e r . c o m