Operating Modes; Triggered Burst; Gated; Sweep - Fluke 291 User Manual

100 ms/s arbitrary waveform generators
Hide thumbs Also See for 291:
Table of Contents

Advertisement

291, 292, 294
Users Manual

Operating modes

Triggered Burst

Each active edge of the trigger signal will produce one burst of the waveform.
Carrier waveforms:
Maximum carrier frequency:
Number of cycles:
Trigger repetition rate:
Trigger signal source:
Trigger start/stop phase:

Gated

Waveform runs while the gate signal is true and stops while false.
Carrier waveforms:
Maximum carrier frequency:
Trigger repetition rate:
Gate signal source:
Gate start/stop phase:

Sweep

Frequency sweep capability is provided for both standard and arbitrary waveforms.
Arbitrary waveforms are expanded or condensed to exactly 4096 points and DDS
techniques are used to perform the sweep.
Carrier waveforms:
Sweep mode:
Sweep direction:
Sweep range:
Sweep time:
Marker:
Sweep trigger source:
Sweep hold:
1-6
Shop for Fluke products online at:
All standard and arbitrary
The smaller of 2.5 MHz or the maximum for the
selected waveform. 100 Msamples/s for arb or sequence.
1 to 1,048,575
0.005 Hz to 100 kHz internal, dc to 1 MHz external.
Internal from keyboard or trigger generator.
External from TRIG IN or remote interface.
±360 °, settable with 0.1 ° resolution, subject to
waveform frequency and type.
All standard and arbitrary.
The smaller of 2.5 MHz or the maximum for the
selected waveform. 100 Msamples/s for arb or sequence.
0.005 Hz to 100 kHz internal, dc to 1 MHz external.
Internal from keyboard, previous channel, next channel
or trigger generator.
External from TRIG IN or remote interface.
±360 °, settable with 0.1 ° resolution, subject to
waveform frequency and type.
All standard and arbitrary except pulse, pulse train and
sequence.
Linear or logarithmic, triggered or continuous.
Up, down, up/down or down/up.
From 1 mHz to 40 MHz in one range. Phase continuous.
Independent setting of the start and stop frequency.
1 ms to 999 s (3 digit resolution).
Variable during sweep.
The sweep may be free run or triggered from the
following sources: manually from keyboard, externally
from TRIG IN input or remote interface.
Sweep can be held and restarted by the HOLD key.