WarmWatch
REPORT INFORMATION
The raw data report downloaded to the USB ash drive by
WarmWatch is written in a comma delimited (CSV) format.
The les are downloaded to a directory identi ed by the
last eight (8) digits of the warmer's serial number. The
report and data les are named by the date of download
and .csv extension. (e.g., 00000000/ 20080101.csv)
The .csv raw data report includes the following categories
for each hourly temperature sample: Entry Number, Year,
Month, Day, Hour, Minute, Second, Warmer Type, Set
point, Control RTD, Logging RTD, Temp Units, Error Bits,
Door, and Warmer State. The .csv raw data report can be
opened in spreadsheet software (such as Microsoft® Excel
or OpenOf ce Calc) for analysis and charting.
An accuracy report indicating how the data was collected
is also provided. The report is a text le with a le name of
the date and an extension of .txt located under the system
serial number folder. (e.g., 00000000/ 20080101.txt)
All temperature samples taken within two (2) hours of an
open door event or within two (2) hours of warmer being
turned on are excluded from the accuracy report.
MM-1068 (MN-29617) • Rev 0 (10/15) • WarmWatch® Combination Blanket/Fluid Warmer • 17
U S B
D O W N L O A D E D
®
R E P O R T S
ACCURACY .TXT REPORT EXAMPLE
20080307.TXT
Fluid Warmer Accuracy Report
Type: Enthermics
Model: 770L
Serial number: 481691
Date of Manufacture: 17-SEP-2007
Date of Report: 07-MAR-2008 13:00:32
Software version: 1.000Final
Frequency of readings: 1.000000 per hour
Period of evaluation: 2008-MAR-05 20:46:45 to
2008-MAR-07 09:05:00
Number of days: 3
Number of readings evaluated: 4118
Number of excluded readings (door open): 242
Mode of warming: injection
Setpoint temperature: 104F
Average temperature: 102.6F
Temperature range: 102.6F -102.7F
Accuracy specifi cation: +0 deg. F., -2 deg. F.
Accuracy evaluation: PASS
Period of evaluation: 2008-MAR-07 09:05:44 to
2008-MAR-07 09:05:44
Number of days: 1
Number of excluded readings (door open): 1
Mode of warming: irrigation
Setpoint temperature: 130F
Too few points exist to generate an accuracy
report for this period of evaluation.
Period of evaluation: 2008-MAR-07 09:06:14 to
2008-MAR-07 12:45:15
Number of days: 1
Number of readings evaluated: 196
Number of excluded readings (door open): 242
Mode of warming: irrigation
Setpoint temperature: 129F
Average temperature: 127.3F
Temperature range: 126.6F -127.7F
Accuracy specifi cation: +0 deg. F., -3 deg. F.
Accuracy evaluation: PASS
Note: The temperature specifi cation is stated for
a warmer in a steady-state condition with the
door closed for at least 2 hours.