Warmwatch® Usb Download Operation - Pedigo WARMWATCH P-2148 Operation & Care Manual

Combination blanket/fluid warming cabinet
Table of Contents

Advertisement

WarmWatch® USB Download Operation
WarmWatch
WarmWatch
uid warmers records cavity temperature data
®
one time every hour. Each recorded data sample includes:
• Time/date stamp
• Current set-point for the chamber
• Air temperature measured by control RTD
• Air temperature measured by data logging RTD
• State of the door (open or closed)
• Warmer type
• Error logging
DOWNLOAD DATA TO A USB FLASH DRIVE
NOTICE: Do not remove the USB ash drive, turn off
warmer, or change warming modes while
downloading data or writing reports to the
USB drive
Recorded data can be downloaded from the warmer at any
desired interval of up to six months between downloads.
(See "Full Memory Information" below.)
A USB ash drive with memory capacity of at least 1 MB is
required. (Not included.) This unit works with most FAT32
formatted USB ash drives.
1. Press ON button if warmer is not already turned
on. Do not turn off or change warming mode while
downloading data.
2. Insert USB ash drive into the USB port located next
to the "INJ" button on the front panel of
the warmer.
3. The main display will show "USB" when it detects
a USB ash drive plugged into unit.
4. The main display will change to "wrt" and the
secondary display will change to "log" prior to
downloading the data.
5. When the data download begins, the main display
will change to "dmP" and the secondary display
will show the percentage of download completed.
The download will take approximately 2 minutes
to complete.
MM-1068 (MN-29617) • Rev 0 (10/15) • WarmWatch® Combination Blanket/Fluid Warmer • 16
U S B
D O W N L O A D
®
C A U T I O N
THIS UNIT IS SENSITIVE TO
ELECTROSTATIC DISCHARGE.
OBSERVE PRECAUTIONS FOR HANDLING
ELECTROSTATIC SENSITIVE DEVICES.
O P E R A T I O N
THERMOSTAT CONTROL AND L.E.D. DISPLAY
MAIN DISPLAY
-
SET
POINT
STATUS
.
.
.
INDICATOR L
E
D
TEMPERATURE
/
ON
OFF
SECONDARY DISPLAY
KEY
ACTUAL TEMPERATURE
6. After the recorded data is downloaded, the main
display will show "wrt" and the second display will
show "rPt". The log report is being downloaded to
the USB drive while these messages are displayed.
7. When the download is complete, the main display
will show "USb" and the second display will show
"donE" until the USB ash drive is removed.
8. After the USB ash drive is removed, the display
returns to normal operation status and the memory
is automatically cleared.
9. Plug USB ash drive into a USB-compatible
computer to access the downloaded report les.
See "Reports" section for more information
about the downloaded reports.
FULL MEMORY INFORMATION
When the data storage memory nears capacity (4,000 data
records or ve months since the last download), the control
will indicate that download is needed by alternately ashing
the actual temperature and the word "Full"on the lower LED
display. Data will continue to be stored until the memory is
full (4,800 records or six months since last download). If the
data is not downloaded before this time, the oldest data will
be overwritten by new data.
Follow steps in "Download Data Via USB Flash Drive" to
download recorded data and clear memory.
STATUS
USB DOWNLOAD
UP
.
.
.
INDICATOR L
E
D
STATUS L
ARROW
OVERTEMP
IRRIGATION
INJECTABLES
DOWN
ALARM KEY
MODE KEY
MODE KEY
ARROW
.
.
.
E
D
USB PORT