B-WIC & B-SMC User Manual
4.3
Software Settings & Configuration
When measuring with B
mode in the Bode Analyzer Suite as shown below:
Note, that you must perform an impedance calibration / correction before you can start measuring.
Check out the next chapter to learn how to perform the Open / Short / Load calibration.
After the calibration has been performed, you can start either a single
sweep or a continuous sweep by clicking on the corresponding icons.
4.4
Calibration / Correction
Before you can start a measurement, Open / Short / Load calibration / correction must be performed.
This moves the reference plane from the Bode 100 front panel to the DUT connection points of the
B
WIC and B
SMC as shown below:
Figure 4-3:
To start a calibration, select either the User-Range Impedance Calibration or the Full-Range Impedance
calibration by clicking on the corresponding icon.
To learn about the difference between Full-Range and User-Range calibration, please refer to the
Bode 100 User Manual.
8
WIC and B
SMC make sure to use the Impedance Adapter measurement
Reference plane after Open / Short / Load calibration
OMICRON Lab