6.5 Other Displays
This will be displayed when the input exceeds the maximum limit of the display range.
˜˜˜˜
This will be displayed when the thermocouple is disconnected.
This will be displayed when the resistance temperature detector is disconnected.
This will be displayed when the input drops below the minimum limit of the display range.
̲̲̲̲
This will be displayed when the resistance temperature detector is shorted.
This will be displayed when the abnormality in the memory element is detected.
ERR0
This will be displayed when the abnormality in the input circuit is detected.
ERR1
This will be displayed when the auto tuning takes more than 3 hours or when error has
ERR2
occurred while the auto tuning is in progress.
This will be displayed when the user tries to change the parameter while the key is locked.
LOC
This will be displayed alternately with the measurement value while the auto tuning process is
in progress.
AT
This will be displayed when the user tries to change the control mode while the F key is being
assigned to RUN/READY.
FUNC
This will be displayed when the user tries to change the control mode while the timer is being
used.
TIME