A notification is displayed in the status field if the controller has been found. Then select the current firmware
file ("xxx.hex") from the directory into which the ZIP file was extracted (see arrow).
After the conversion, tap the button "Update firmware" to start the firmware upload:
The program can be ended after the successful firmware update by closing the window.
The USB A-B cable can now be disconnected from the controller.
Step 3: Reset DIP switch S3
After performing a firmware update, only the coding switch 9 of the DIP switch S3 must be reset to ON:
Risk of controller failure! The controller only accepts the settings for DIP switch S3 after the
controller has been briefly disconnected from the power supply!
After the firmware update, the saved settings can be reloaded into the case controller from LDSWin as
required..
Firmware V3.00
05.08.2019
52/100