600 V CAT IV 120/ 230/ 400/ 208V 400V 690V Connector sockets for phases L1 ... L3 Rotary LEDs for displaying frequency and phase sequence Phase sequence and frequency Green arrow: Clockwise rotation, LEDs light up green and rotate clockwise Momentary frequency: Indicated by LED which briefly lights up red Dashed red arrow: Counterclockwise rotation,...
Page 3
Meanings of Symbols on the Device Warning concerning a point of danger (attention: observe documentation) Ground Continuous, doubled or reinforced insulation CAT IV Maximum allowable voltage between the connector sockets (1) and ground it is 600 V category IV. EU conformity marking This device may not be disposed of with house- hold trash.
Table of Contents Page Safety Instructions ............5 Applications ..............7 Intended Use / Use for Intended Purpose ......7 Use for Other than Intended Purpose ........7 Liability and Guarantee ............7 Opening the Instrument / Repairs ........7 Initial Startup ..............
Safety Instructions Observe this documentation, in particular all included safety information, in order to protect yourself and others from injury, and to prevent damage to the instrument. • Carefully and completely read and adhere to these operating instructions. This document can also be accessed at http:// www.gossenmetrawatt.com.
Page 6
• The instrument must not be exposed to direct sunlight. • The instrument and the included accessories may only be used for the measurements described here and in the operating instructions for the instrument. • Only use the instrument and the accessories within the specified measuring category.
Applications Intended Use / Use for Intended Purpose The instrument is a phase sequence indicator with elec- tronic rotary dial, frequency display and nominal line volt- age display. Note! It’s not a voltmeter in the usual sense of the word – it only indicates voltage within the specified ranges.
Initial Startup Batteries The phase sequence indicator is shipped ready to use with four AA batteries. Before initial startup, be sure to read section 7.1, “Battery”. Battery Test If none of the LEDs light up after applying voltage, or if they’re very dim, the batteries need to be replaced.
LED indicates the next lower volt- age because in this case the instrument no longer has an internal reference point. Attention! The L1 socket on the METRAPHASE 1 must always be contacted first in order to prevent erroneous measurements. 3-Pole Connection...
Page 10
Phase sequence can also be determined with just 2 con- nector cables if line frequency is 50 or 60 Hz: Contact phase L1 first via connector socket L1 and then connect phase L2 via connector socket L2 (see figure A). Phase LEDs L1 and L2, as well as the corresponding voltage value LED, light up red.
Characteristic Values, Phase Sequence Indicator Line Frequency Nominal Display Range Tolerance Range Display Frequency 50 Hz 50 Hz 49.4 ... 50.7 Hz 49.2 ... 50.9 Hz – LED active 50 Hz and60 Hz — 51.1 ... 57.9 Hz 50.9 ... 58.1 Hz –...
Page 12
Reference Conditions +23 C 2 K Ambient temperature Relative humidity 40 ... 60% 4.5 V 0.25 V Battery voltage 230/400 V 0.5% Line voltage 50 Hz 0.1 Hz Line Frequency Line voltage waveform Sine, deviation between TRMS and rectified value < 1% Ambient Conditions –25 C ...
DIN EN 61326-1 Electrical equipment for measurement, control and labora- VDE 0843-20-1 tory use – EMC requirements – Part 1: General require- ments CE Declaration The instrument fulfills all requirements of applicable EU directives and national regulations. We confirm this with the CE mark.
Housing No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives or solvents. Returns and Environmentally Sound Disposal The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German electrical and electronic device law).
Repair and Replacement Parts Service Calibration Center and Rental Instrument Service If required please contact: GMC-I Service GmbH Service Center Beuthener Str. 41 D-90471 Nürnberg, Germany Phone: +49-911-817718-0 Fax: +49-911-817718-253 e-mail service@gossenmetrawatt.com www.gmci-service.com This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.