Figure 2. Contact the supplier if any items are missing. ATTENTION: The included USB driver CD must be installed before connecting the AEROCET 531S USB port to your computer. If the supplied drivers are not installed first, Windows may install generic drivers that are not compatible with this product.
2.2. Layout The following figure shows the layout of the AEROCET 531S and provides a description of each of the components. Inlet Nozzle AT/RH Connector USB Port Serial Port Power Switch Charging Jack Figure 3 – AEROCET 531S Layout Component...
RS-232 2.4. Initial Operation Before operating the AEROCET 531S for the first time, it is recommended that the unit be fully charged. Information regarding charging the battery is found in section 6 of this manual. Complete the following steps to verify proper operation.
4.3. Power Up AEROCET 531S power is controlled by a slide switch located on the right hand side of the unit. Move the power switch to the on position (towards the top of the case) to power up the unit.
3 lines scroll to display the full list. AT and RH data will precede the location number when the AT/RH probe is attached. Ambient temperature (AT) can be displayed in units of Celsius (C) or Fahrenheit (F). AEROCET 531S Manual Page 9 AEROCET 531S-9800 Rev A.docx...
0.0 ug Figure 7 – Operate Screen Sampling 4.4.2. Sample Status The top line of the Operate screen displays the status of the AEROCET 531S while the unit is sampling. The following table shows the various status messages: Status Description AEROCET 531S is starting the sample and is waiting for the STARTING...
5.2. 4.5.5. Real-Time Serial Report AEROCET 531S provides real-time report out the serial port at the end of each sample. The format of the report is controlled by the REPORT setting (Section 5.4). Section 7.2 covers the real-time output format.
Press to select the item. Press ▲ or ▼ to change the ENTER setting. Press to save the setting or to cancel and return to the main value. ENTER AEROCET 531S Manual Page 12 AEROCET 531S-9800 Rev A.docx...
HOLD time is 60 seconds or less. The pump will stop after each sample, and start a few seconds before the next sample, if the HOLD time is greater than 60 seconds. HOLD times greater than 60 seconds will increase pump life. AEROCET 531S Manual Page 13 AEROCET 531S-9800 Rev A.docx...
Figure 12 – SETTINGS Screen 5.3.1. VOLUME The AEROCET 531S supports total counts (TC), particles per liter (/L), particles per cubic foot (CF), and particles per cubic meter (M3). Particle count information updates while the unit is sampling. Concentration values (/L, CF, M3) are time dependent so these values may fluctuate early in the sample;...
SERIAL: RS-232 Figure 13 – SERIAL Screen 5.4.1. REPORT setting determines the behavior of the AEROCET 531S serial output and REPORT report format. The settings are NONE, CSV or PRINTER. The following table lists the settings and describes their meanings. See section 7 for report examples.
000 prints all locations. Settable from 0 – 999. The date/time to begin printing sample events from. 01 JAN’00 00:00 The date/time to stop print samples at. 18 AUG’06 13:23 AEROCET 531S Manual Page 16 AEROCET 531S-9800 Rev A.docx...
5.7. MEMORY Screen The AEROCET 531S memory is composed of a single file which contains the data from sample events. Every time a sample is completed, the AEROCET 531S stores that data into the memory. The AEROCET 531S memory is circular, meaning when the memory is full, the unit will start overwriting the oldest saved samples with new samples.
5.8. SET FLOW Screen The AEROCET 531S has a factory calibrated flow rate of 0.1 CFM (2.83 LPM). Use the following procedure to adjust the flow rate when a periodic flow rate check (Section 8.3) indicates a flow rate error greater than +/- 5%.
To charge the battery, connect the battery charger to an AC power outlet and the DC barrel connector to the socket on the right side of the AEROCET 531S. The battery charger is universal and will work with power line voltages of 100 to 240 volts, 50 to 60 Hz.
AEROCET 531S. ATTENTION: The included USB driver CD must be installed before connecting the AEROCET 531S USB port to your computer. If the supplied drivers are not installed first, Windows may install generic drivers that are not compatible with this product..
7.1. Commands AEROCET 531S provides serial commands for accessing stored data and settings. All commands are terminated by a carriage return. Also, these commands are not case sensitive. The following table lists the available commands. These commands are available via USB, RS-232 and RS-485 hardware interfaces. The settings (baud rate, parity and stop bits) must match the computer setting for proper communication regardless of the hardware interface type (USB, RS-232 or RS-485).
The real time output occurs when the unit finishes a sample. The output format is either off (NONE), a comma separated value (CSV) or a printer (PRINTER) depending on the REPORT style settings (see section 5.4.3) AEROCET 531S Manual Page 22 AEROCET 531S-9800 Rev A.docx...
4. The sample time is fixed at 60 seconds. Low battery 5. Status bits combinations are possible. See the Status Bits table. Sensor error For example, 48 (00110000B) = Low battery and Sensor error. AEROCET 531S Manual Page 23 AEROCET 531S-9800 Rev A.docx...
4. The sample time is fixed at 60 seconds. Low battery 5. Status bits combinations are possible. See the Status Bits table. Sensor error For example, 48 (00110000B) = Low battery and Sensor error. AEROCET 531S Manual Page 24 AEROCET 531S-9800 Rev A.docx...
Due to the nature of the instrument, there are no customer serviceable components in the AEROCET 531S. The case of the AEROCET 531S should never be removed or opened for any reason. Opening or removing the case of the AEROCET 531S voids the warranty and may result in exposure to laser radiation, which can cause eye injury.
8.4. Annual Calibration The AEROCET 531S should be sent back to Met One Instruments yearly for calibration and inspection. The annual calibration cannot be performed by the customer because this calibration requires specialized equipment and a skilled technician. Met One Instruments maintains a calibration facility for calibrating particle counters according to industry accepted methods such as ISO, JIS and NIST.
The AEROCET 531S case should never be removed or opened for any reason. Opening or removing the case will void the warranty and may result in exposure to laser radiation, which can cause eye injury.
Any product found to be defective during the warranty period will, at the option of Met One Instruments, Inc., be replaced or repaired. In no case shall the liability of Met One Instruments, Inc. exceed the purchase price of the product.