Quarter-Bridge Completion and Shunt Calibration

The RM-4339 provides software selectable quarter-bridge completion resistors that enable the
NI PXIe-4339 to be used with quarter-bridge strain gages. You can select 120 Ω, 350 Ω or 1 kΩ
completion resistors.
Quarter-bridge shunt calibration resistors are also provided on the RM-4339. These resistors are
labeled SCB_50k and SCB_100k in the switch circuit diagram of Figure 8. There are two
different shunt calibration resistance values that you can select; 50 kΩ and 100 kΩ. The
quarter-bridge shunt calibration resistors are dedicated to the quarter-bridge completion resistors
and are selectable through NI-DAQmx. NI recommends using the 50 kΩ shunt calibration
resistor for shunt calibration of 120 Ω quarter-bridge strain gages and using the 100 kΩ shunt
calibration resistor for 350 Ω and 1 kΩ quarter-bridge strain gages.
If quarter-bridge completion is not used, the QTR/RS- pin of the RJ50 connector is connected to
the RS- pin of the NI PXIe-4339. Refer to the NI PXIe-4339 User Manual for more information
about quarter-bridge measurements.
You can access SCB_50k and SCB_100k using NI-DAQmx attributes/properties.
LabVIEW property names and locations:
AI.Bridge.ShuntCal.ShuntCalBResistance
Location: AI»General Properties»Signal Conditioning»Bridge»ShuntCal»Shunt Cal
B Resistor Value
AI.Bridge.ShuntCal.ShuntCalBActualResistance
Location: AI»General Properties»Signal Conditioning»Bridge»Shunt Cal B Actual
Resistor Value

Shunt Resistors

The RM-4339 provides built-in 100 kΩ shunt resistors that can be programmatically connected
to the SCA terminal pairs on the RJ50 connectors. In addition you can use the optional
SCAL-4339 Shunt Calibration Accessory to provide user provided shunt calibration resistors to
the SCA terminals of each channel. Refer to the
Refer to Table 7 for signal names and descriptions.
You can choose between built-in 100 kΩ resistors and external user provided resistors using an
NI-DAQmx channel attribute/property.
LabVIEW property names and locations:
AI.Bridge.ShuntCal.ShuntCalASrc
Location: AI»General Properties»Signal Conditioning»Bridge»ShuntCal»Shunt Cal
A Source
Possible values: BuiltIn or UserProvided
22 | ni.com | NI PXIe-4339 and RM-4339 User Guide and Terminal Block Specifications