CALIBRATION PROCEDURE
NI PXIe-4463
DSA Analog Output
This document contains the verification and adjustment procedures for the National Instruments
PXIe-4463 with either BNC or mini-XLR connectors. For more information about calibration
solutions, visit
ni.com/calibration
Contents
Software.................................................................................................................................... 2
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 5
Initial Setup............................................................................................................................... 6
Self-Calibration ................................................................................................................ 6
Verification ............................................................................................................................... 6
AO Noise Verification.............................................................................................. 13
Adjustment................................................................................................................................ 18
Timebase Adjustment ....................................................................................................... 19
EEPROM Update ..................................................................................................................... 20
Reverification ........................................................................................................................... 20
.