CALIBRATION PROCEDURE
NI PXIe-4464
4-Channel, 204.8 kS/s DSA Input
This document contains the verification and adjustment procedures for the National Instruments
PXIe-4464. The NI PXIe-4464 is available in two input connector variants: BNC and mXLR.
The calibration procedure is common between them; only the connecting cabling is different. In
this document, the BNC and mXLR variants are referred to inclusively as the NI PXIe-4464. For
more information about calibration solutions, visit
.
ni.com/calibration
Contents
Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 2
Test Conditions......................................................................................................................... 5
Initial Setup............................................................................................................................... 5
Verification ............................................................................................................................... 6
Adjustment................................................................................................................................ 16
Reverification ........................................................................................................................... 18
Updating the Calibration Date without Performing Adjustment.............................................. 18
Specifications............................................................................................................................ 19
Where to Go for Support .......................................................................................................... 19

Software

Calibrating the NI PXIe-4464 requires the installation of NI-DAQmx on the calibration system.
NI-DAQmx 14.1 added driver support for the NI PXIe-4464. For the list of devices supported
by a specific release, refer to the NI-DAQmx Readme, available on the version-specific
download page or installation media.
You can download NI-DAQmx from
. NI-DAQmx supports LabVIEW,
ni.com/downloads
LabWindows
/CVI
, C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx, you
only need to install support for the application software that you intend to use.

Advertisement

Table of Contents
loading

Summary of Contents for National Instruments NI PXIe-4464

  • Page 1: Table Of Contents

    PXIe-4464. The NI PXIe-4464 is available in two input connector variants: BNC and mXLR. The calibration procedure is common between them; only the connecting cabling is different. In this document, the BNC and mXLR variants are referred to inclusively as the NI PXIe-4464. For more information about calibration solutions, visit ni.com/calibration...
  • Page 2: Documentation

    Documentation Consult the following documents for information about the NI PXIe-4464, NI-DAQmx, and your application software. All documents are available on and help files install with the ni.com software. NI PXIe-4464 Getting Started Guide NI-DAQmx driver software installation and hardware setup...
  • Page 3 NI 156789-XX or (quantity 4) NI 140150-XX 0R460.46 m 0R910.91 m 2R42.4 m — BNC F-F Adapter Pomona Electronics AI Offset (quantity 4) 3283 Required only for the mXLR variant. NI PXIe-4464 Calibration Procedure | © National Instruments | 3...
  • Page 4 Connecting the NI PXIe-4464 The NI PXIe-4464 provides connections for four channels. Figure 1 shows the front panel connections of the NI PXIe-4464 with BNC connectors and the NI PXIe-4464 with mXLR connectors. Figure 1. NI PXIe-4464 BNC and mXLR Connector Front Panel...
  • Page 5: Test Conditions

    The following setup and environmental conditions are required to ensure the NI PXIe-4464 meets published specifications. • Keep connections to the NI PXIe-4464 as short as possible. Long cables and wires act as antennas, picking up extra noise that can affect measurements. •...
  • Page 6: Verification

    Analog Input Performance Verification This section verifies the analog input (AI) performance of the NI PXIe-4464. Some, but not all, performance characteristics are affected by adjustment and are subject to reverification after adjustment. Characteristics affected by adjustment have As Left test limits in addition to As Found test limits.
  • Page 7 -10, 0, 10, 20, and 30 as required. Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. NI PXIe-4464 Calibration Procedure | © National Instruments | 7...
  • Page 8 Finite number of samples Rate 204800.0 Samples per channel 102400 Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. 8 | ni.com | NI PXIe-4464 Calibration Procedure...
  • Page 9 Consequently, the verification of AC-coupled gain flatness covers both configurations. Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. NI PXIe-4464 Calibration Procedure | © National Instruments | 9...
  • Page 10 Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. Start the task. Acquire readings with the NI PXIe-4464. 10 | ni.com | NI PXIe-4464 Calibration Procedure...
  • Page 11 1.01158 * V 1kHz 1kHz 1kHz 1kHz Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. NI PXIe-4464 Calibration Procedure | © National Instruments | 11...
  • Page 12 -20 dB gain setting at the 204.8 kS/s sample rate. Table 11. AI Noise Limits = 51.2 kS/s = 204.8 kS/s Device Gain (dB) Max (µV Max (µV 14.0 32.9 12 | ni.com | NI PXIe-4464 Calibration Procedure...
  • Page 13 Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module Refer to Table 3 for the maximum and minimum voltages to specify to ensure the specified gain range is used on the module. NI PXIe-4464 Calibration Procedure | © National Instruments | 13...
  • Page 14 Start the task. Acquire readings with the NI PXIe-4464. The readings are not used; the task is simply run to set the device configuration. Measure the output current on the digital multimeter. Stop and clear the task. Compare this value to the appropriate limit in Table 13.
  • Page 15 Table 15. Timebase Frequency Accuracy Limits Function Generator Output As Found Test Limit As Left Test Limit Amplitude Frequency (kHz) (kHz) (kHz) (kHz) (kHz) 10.000 9.99973 10.00027 9.99992 10.00008 Stop and clear the task. NI PXIe-4464 Calibration Procedure | © National Instruments | 15...
  • Page 16: Adjustment

    Adjustment If the NI PXIe-4464 successfully passed each of the verification procedures within the As Left test limits without adjustment, then an adjustment is recommended, but not required, to warrant the published specifications for the next calibration interval. If the device was not within the As Left test limits, you can perform the adjustment procedure to improve the device accuracy and reset the calibration interval.
  • Page 17 ) signal, no offset, at a frequency of 90.000 kHz. 17. Call (DAQmx Adjust DSA Timebase Calibration VI) DAQmxAdjustDSATimebaseCal with the following parameters to perform the timebase adjustment: calHandle: myCalHandle referenceFrequency: 90000.0 NI PXIe-4464 Calibration Procedure | © National Instruments | 17...
  • Page 18: Reverification

    You can update the calibration date without performing the adjustment procedure. This might be appropriate if the NI PXIe-4464 successfully passed each of the verification procedures within the As Left test limits without adjustment. Complete the following steps to update the...
  • Page 19: Specifications

    NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS.