Setup commands
Command
Description
Write/Read
SCN=/SCN=?
Set the comparator set number.
Command/
12SCN=3
Return format
3= 1 : Set number 1
2 : Set number 2
3 : Set number 3
4 : Set number 4
REF=/REF=?
Set whether the reference point is to be used or not.
Command/
12REF=3
Return format
3= 0 : Reference point used
(reference point load mode established)
1 : Reference point not used
(counter mode established)
LCLR
Clear the reference point setting
Command
(reference point offset value)
format
12LCLR
STTERM=/STTERM=? Select the external start input function.
Command/
1STTERM=3
Return format
3= 0 : Start function
1 : Hold (latch) function
RSSEP=/RSSEP=? Select the RS-232C data transfer format
Command/
(separator).
Return format
1RSSEP=3
3= 0 : SPACE
1 : "
C R
RSFORM=/RSFORM=? Select the RS-232C output data format.
Command/
1RSFORM=3
Return format
3= 0 : Mode 1
1 : Mode 2
2 : Mode 3
(For details on modes 1 to 3, see section 6-1-3.)
MG10-P1 / MG10-P2
" + "
L F
"
Command
Description
Write/Read
RSTRG=/RSTRG=? Select the RS-232C trigger or internal timer.
Command/
1RSTRG=3
Return format
3= 0 : RS-232C trigger low-speed mode,
internal timer not used
1 : RS-232C trigger high-speed mode,
internal timer not used
2 : RS-232C trigger not used, internal timer 0.2 s.
3 : RS-232C trigger not used, internal timer 0.5 s.
4 : RS-232C trigger not used, internal timer 1.0 s.
5 : RS-232C trigger not used, internal timer 5.0 s.
6 : RS-232C trigger not used, internal timer 10 s.
7 : RS-232C trigger not used, internal timer 30 s.
8 : RS-232C trigger not used, internal timer 60 s.
9 : RS-232C trigger not used, internal timer 300 s.
Low-speed mode : Mechanical contact input
High-speed mode: Electronic contact input
n
The internal timer cannot be used when using a link connection.
Be absolutely sure that all units have been set to either RS-TRG
low-speed mode or high-speed mode. Normal communications
are impossible when the settings for each unit differ.
SCALE
Set the scale unit (mm or inch).
Command
1SCALE=3
format
3= 0 : mm
1 : 1/25.4 mm
CLOSE
Close the parameter setup.
Command
CLOSE
Close the parameter setup and store the setup data.
(relay, switch, etc.)
(transistor, etc.)
25
(E)