Properly Aligning the Charger

1
Press the power button. The indicator light will come on again, and the Charger will begin
beeping repeatedly to signal that it is searching for the Stimulator.
2
Position the Charger over the Stimulator.
The beeping will stop when the Charger is partially aligned with the Stimulator. For
optimal charging, make sure the Charger is centered over the Stimulator.
Charging time may increase if the Charger is not fully aligned with the Stimulator. To
fully align the Charger over the Stimulator, position the bottom curve of the Charger
approximately 1 cm lower than the bottom curve of the Stimulator, as shown.
If you are having trouble properly aligning the Charger or experiencing longer than
normal charging times, see section Tips for Properly Aligning the Charger.
3
Secure the Charger over the Stimulator either by pressing the adhesive to the skin over the Stimulator, or letting go
of the Charging Collar.
Note: If the Adhesive Patch is accidentally located in the wrong place, or if the Charging Collar moves out of alignment,
the Charger will start beeping again. Use a new Adhesive Patch or readjust the Charging Collar to place the
Charger back into the fully aligned position.