Download Print this page

Imada MX-500N Series Instruction Manual page 16

Test stand

Advertisement

Instruction Manual MX-500N / MV-500NII Series
MX-500N series has three operation modes. This unit has "Overload stop" feature that
always works while the gauge is connected by optional cable and it is powered on. Also
"Force control" feature works when the gauge is connected.
The cable to connect between gauge and the test stand is sold separately. Some force
gauge is not compatible with those features
Manual mode
Single cycle mode
Continuous cycle mode
Additional mode effective when the gauge is connected
Force control mode
3. It is in Manual mode when you turn the test stand on.
4. Every pressing
Continuous cycle with LED indication.
Operating MX series
Three modes
Crosshead moves while pressing
Crosshead starts by
direction when it reaches travel limit at its maximum speed. It
automatically stops at another travel limit.
Continuously goes up and down between its upper and lower
travel limit.
Connecting with the gauge, the crosshead moves according
to the comparator signals (OK/NG) from the gauge.
Stops when above signal shows +OK.
Automatically control the crosshead to keep the
force between high and low comparators setup
values.
Selecting modes`
selects mode in the sequence of Manual - Single cycle -
16
. It will reverse its travel
Manual No. MXV50801E

Advertisement