SONOLINE G20 System Delivery Procedure
6. Verify that the system boots to a blank screen.
Fig. 10: Blank Screen
NOTE
Siemens
Medical Solutions USA, Inc.
The boot sequence takes approximately 80 seconds, 150 seconds
if POST is set to run during the boot sequence. If the system does
not boot within 80 to 150 seconds, press STANDBY and wait for
the system to shut down. Verify that the power connections are
correct, and then boot the system. If system does not boot, report
this to the Siemens Uptime Service Center.
US06-101.812.01.02.02
10.04
Page 13 of 26