Power On the System

1. Attach the power cord safety lock to the system.
- Verify that the MAINS power switch is off.
- Choose the correct power cord and safety lock for the system (100V/115V or 230V).
NOTE
- Attach the safety lock to the system power inlet as shown in the figure below. Insert
the lower post of the safety lock first then the upper post.
Fig. 6:
Pos. A
Pos. B
Pos. C
Pos. D
- Position the safety lock on the left side of the power inlet; plug the power cord into the
power inlet.
- Press the safety lock on to the power cord; the safety lock must snap into the locking
position.
2. Plug the other end of the AC power cord into an approved power source.
SONOLINE G20
SONOLINE G20 System Delivery Procedure
Two power cords and safety locks are provided; use only the ap-
propriate power cord and safety lock for the system.
For 100 and 115V systems, use the white power cord (part num-
ber KP4819Y31AY3) with the small safety lock (part number
4H540092-0DB).
For 230 V systems, use the gray power cord (part number
HG14A13V3.5M) with the large safety lock (part number
4H540070-0DB).
Power cord Safety Lock
Safety lock
Power inlet
Lower post of the safety lock
Upper post of the safety lock
US06-101.812.01.02.02
10.04