ABB PGS100 Operating Instruction page 25

Gauge and absolute pressure transmitters
Table of Contents

Advertisement

P G S 1 0 0 A N D PA S 1 0 0 | P R E S S U R E T R A N S M I T T E R S | O I/ P G S/ PA S 1 0 0 - E N R E V. A
HMI -
Code
Error
NAMUR
Code
M
xxx.003
Electronics
temperature
failure
M
026.024
Electronics
NV-memory
burn error
M
028.018
Sensor
NV-Memory
burn error
M
024.036
Power
supply
warning
M
022.041
Electronics
temperature
out of limits
Description
The measurement accuracy could be
decreased because the digital to analog
converter (DAC) remains compensated
with the last valid value of the electronic
temperature.
Permanent storing of modified data
into the electronic Non-Volatile memory
has not been successfully executed.
The device continues to work without
problems but, after the next power cycle
there is the risk to lose the last used
device setting/configuration.
Permanent storing of modified data
into the Sensor Non-Volatile memory has
not been successfully executed.
The device continues to work without
problems but, after the next power cycle
there is the risk to lose the last used
device setting/configuration.
The device power supply has been
detected too close to its higher or lower
acceptable limits.
The electronics temperature is working
beyond its High or Low specification
Possible cause
The circuitry for the sampling
of the electronics temperature has
• The device could have been
powered off immediately
after the modification of the
data that was not fully written
in the NV-memory.
• The NV-memory has reached
its limit of maximum number
of allowed writings
• Generic defect of the memory
component.
The device could have
been powered off immediately
after the modification of the data
that was not fully written in the
NV-memory.
The NV memory has
reached its limit of maximum
number of allowed writings
Generic defect of the
memory component
If the power supply is close to the
highest limit very probably it is due
to a not correct external power
supply while if the power supply is
close to the lowest limit, it could
be due to a defective or not
adequate external power supply or
anomalous voltage drop on the
loop i.e. too long or not correct
This condition could be due to the
circuitry for the sampling of the
limits.
temperature that has failed or due
to the transmitter installation e.g.
steam leakage directed on to
transmitter housing, or critical
ambiental conditions.
Suggested action
The device should be
replaced as soon as
failed.
possible.
The device should be
replaced as soon as
possible.
Retry to write the same
configuration values
but, if the error
persists, consider to
replace the device as
soon as possible
Check the voltage at the
terminal block and if it
is not within the valid
range check the
external power supply
or the loop.
wires.
If necessary, use a HART
configurator in order to
know in detail of which
limit (high or low) the
electronics temperature
is beyond.
Check the transmitter
installation to identify
possible cause of the
abnormal measured
temperature and take
possible actions to
minimize the
temperature impact/
influence like to insulate
the device from the
external temperature
influence.
25
Tx
Response
No effect
No effect
No effect
No effect
No effect