Keysight 42030A Operation And Service Manual

Keysight 42030A Operation And Service Manual

Four-terminal pair standard resistor set

Advertisement

Keysight 42030A Four-Terminal Pair
Standard Resistor Set
Operation and
Service Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 42030A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight 42030A

  • Page 1 Keysight 42030A Four-Terminal Pair Standard Resistor Set Operation and Service Manual...
  • Page 2 Keysight Declarations of Conformity for this 2.101, pursuant to FAR 12.211 and Technologies, Inc. as governed by product and for other Keysight prod- 27.404.2 and DFARS 227.7102, the United States and international ucts may be downloaded from the U.S.
  • Page 3: Table Of Contents

    Non-Keysight Packing Materials ........
  • Page 4 Contents Keysight 42030A Four-Terminal Pair Standard Resistor Set...
  • Page 5: General Information

    Keysight 42030A Four-Terminal Pair Standard Resistor Set Operation and Service Manual General Information Introduction This chapter provides the 42030A Four-Terminal Pair Standard Resistor Set description, specifications, and related general information. Description The 42030A Four-Terminal Pair Standard Resistor Set consists of nine precision resistor standards.
  • Page 6: Initial Inspection

    Keysight Part No. Not Assigned Calibration Report The 42030A is designed for use in calibrating the Keysight LCR meters which have four-terminal pair measurement terminals, and can be directly connected to the measurement terminals. The LCR meter documentation will give detailed information on using the 42030A.
  • Page 7: Repackaging For Shipment

    The following general instructions should be used for re-packing with commercially available packing materials: 1. If shipping to Keysight Technologies office or service center, attach a tag indicating the type of service required, return address, model number, and full serial number.
  • Page 8: Specifications

    General Information Specifications Specifications This section gives the complete specifications for the 42030A Four-Terminal Pair Standard Resistor Set. When the 42030A is shipped from the factory, it meets the specifications listed in this section. DC Resistance Table 1-2 lists the nominal DC resistance for each resistance module of the 42030A.
  • Page 9: Operating Conditions

    General Information Operating Conditions Operating Conditions The 42030A must be operated under the ambient environmental conditions listed in Table 1-3. Table 1-3 Operating Conditions Temperature 23 °C ± 5 °C Relative Humid ity ~70% Dimensions Table 1-4 shows the 42030A dimensions.
  • Page 10: Supplemental Performance Characteristics

    1. Under temperature condition: 23 °C ± 5 °C 2. Under environmental conditions of: Temperature 23 °C ± 5 °C, Relative Humidity ~70% Storage Conditions The 42030A must be stored or shipped under the ambient environmental conditions listed in Table 1-7.
  • Page 11: Calibration And Repair

    Keysight 42030A Four-Terminal Pair Standard Resistor Set Operation and Service Manual 2 Calibration and Repair Introduction This chapter provides the calibration and repair information for the 42030A Four-Terminal Pair Standard Resistor Set. Calibration The 42030A is calibrated in the parameters given in...
  • Page 12 Calibration Uncertainty Frequency Series Resistance Reactance 1 MHz ± 0.12% ± 0.10 Ω 2 MHz ± 0.12% ± 0.10 Ω 3 MHz ± 0.15% ± 0.15 Ω 4 MHz ± 0.15% ± 0.15 Ω Keysight 42030A Four-Terminal Pair Standard Resistor Set...
  • Page 13 ± 0.08% ± 0.20 µS Table 2-7 42039A 100 kΩ Resistor AC Calibration Uncertainty Calibration Uncertainty Frequency Parallel Resistance Susceptance 100 kHz ± 0.08% ± 0.02 µS 1 MHz ± 0.12% ± 0.15 µS Keysight 42030A Four-Terminal Pair Standard Resistor Set...
  • Page 14: Calibration Interval

    Calibration and Repair Repair The 42030A must be calibrated at a facility which satisfies the calibration uncertainties given in Table 2-2 through Table 2-7. Keysight Technologies' calibration laboratories satisfy all of these calibration uncertainties. For complete information on calibration, contact your nearest Keysight Technologies sales and service office.
  • Page 15 This information is subject to change without notice. © Keysight Technologies 1992 - 2019 Edition 3, August 28, 2019 *42030-90001* 42030-90001 www.keysight.com...

Table of Contents