目錄 1. 簡介.........1 1.1 內容物簡介.......1 1.2 外觀說明........2 1.3 產品特色........3 1.4 產 品 使 用 注 意 事 項 ....4 2. 基本使用........5 2.1 開機/關機........5 2.2 M i c r o S D 卡安...
Introduction 1.1 Description of Contents The shipment should contain the following objects. If anything is missing, please contact the distributor. Transmission RM785 Suitcase Line Charger Extraction Bottle Warranty Card Manual ~ 23 ~...
1.3 Product Features 1. Built-in 785nm laser. 2. Broad Raman shift range (400cm ~ 3000cm 3. Support SERS measurement. 4. System built-in Raman spectrum database, users can expand the database by themselves. 5. Provide data input, output and comparison function. 6.
2. To assure measurement accuracy, we suggest send the unit back to us for calibration once per year. 3. RM785 uses lithium battery. Please do not store or operate it under high temperature condition to prevent the battery from combustion or explosion.
2 Basic Uses 2.1 Power On/Power Off Power On: Short press the power button, green light will go on. Power Off:Long press the power button for 3 seconds, green light will go off. Diagram for reference: Power on will have the green light on. Power off will have the green light off.
2.2 Micro SD Card Installation Method Install Micro SD card: Please follow the instruction to insert Micro SD card. Take out Micro SD card: Press the card Take out the card ~ 28 ~...
2.3 Measurement Method Process flow: 1. Put liquid sample into the extraction bottle. 2. Tighten the extraction bottle cap. 3. Put extraction bottle into the sample holder. 4. Put on the sample holder cap and tighten it. 5. Press the measurement button and start measurement.
System Reset When the system is down, please press reset button to reset the system. Diagram for reference: # Suggest to use pencil tip pressing to reset the system ~ 30 ~...
2.5 Battery Charging Mode 1. Connect charger to RM785 charging hole to start charging. 2. For charging when power is off, power button will show red light. After charging is completed, red light will go off. 3. For charging when power is on, the upper right corner of screen will show a battery logo and the power button will show red light.
2.6 USB Mode 1. RM785 supports USB. Users can select storage device mode or PC connection mode. 2. Storage device mode (MSC): access micro SD. 3. PC connection mode (SPD): use Opticker program to operate RM785 measurement from Diagram for reference:...
3.1 MATCH Measurement Description Description: compare Raman signal to database, identify material composition. MATCH interface: ~ 34 ~...
Page 38
1. System mode. 2. System status. 3. Battery power level. 4. Measurement result (Raman spectrum). 5. Comparison function -- after measurement press the button to indicate whether the sample is a substance in the database. If the database does not have the substance, it is not identified, as in the following figure.
3.2 COMP Measurement Description Description: Simultaneously show several measurement results on the same coordinate axis for user comparison. COMP interface: 1. System mode. 2. System status. 3. Battery power level. 4. Measurement result (Raman spectrum). 5. Record measurement spectrum. 6. Parameter setup. 7.
3.5 SYSTEM Operating Instructions Description: Users can set up system parameters by themselves SYSTEM interface: 1. Sound setup (ON, OFF). 2. Backlight setup (30% ~ 100%). 3. Automatic power off time setup (0 ~ 30 minutes). 4. USB mode setup (MSC, SPD). 5.
5. Limited Product Warranty strider tech guarantees that RM785 products will provide consumers with free repair or replacement for the products of the same grade during the warranty period, such as defects or malfunctions in materials or functions. This warranty does not cover any damage resulting from improper...
Page 45
5.1 Product within Warranty Period 1. RM785 has 2 years of original manufacturer warranty. 2. Consumables for the product are not covered by the warranty. 3. New product replacement: Consumers within 7 days after the purchase of a product that has functional defects or failure, can apply for a replacement.
Page 46
Opticker Download And Install twinno 1. Connect to official website: www.twinno.com Click Install file 4. Download file and install Opticker_Install.exe. 5. After install, Opticker.exe path appear on desktop. 6. Execute Opticker.exe start up the program. ~ 43 ~...
Need help?
Do you have a question about the RM785 and is the answer not in the manual?
Questions and answers