5240
5 100 - 5 800
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on
RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum
separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation:
E=6/d √P
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the
IMMUNITY TEST LEVEL in V/m.
100275_MYO-TENS_USER MANUAL_REV.00.indd 53
WLAN 802.11 a/n
Pulse modulation b)
217 Hz
0,2
0,3
9
53
EN
08/08/2019 15:15:03