Tab. 2) Guidance and manufacturer's declaration - electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the
Device should assure that it is used in such an environment
Immunity test
IEC 60601 Test level
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
Power supply
transient/burst
lines :±2 kV
IEC 61000-4-4
input/output lines:±1 kV
Surge
line(s) to line(s): ±1 kV
IEC 61000-4-5
line(s) to earth: ±2 kV
100 kHz repetition frequency
100275_MYO-TENS_USER MANUAL_REV.00.indd 47
Compliance level
Electromagnetic
environment - guidance
±8 kV contact
Floors should be wood,
±15 kV air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %
Not applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
Not applicable
Mains power quality
should be that of a typical
commercial or hospital
environment.
47