Tab. 3) Guidance and manufacturer's declaration - electromagnetic immunity
The customer or the user of the Device should assure that it is used in such an environment
IEC 60601
Immunity test
Test level
Conduced RF
150KHz to 80MHz:
IEC61000-4-6
3Vrms
6Vrms (in ISM and
amateur radio
bands)
80% Am at 1kHz
100275_MYO-TENS_USER MANUAL_REV.00.indd 49
The Device is intended for use in the electromagnetic environment specified below.
Compliance level
Electromagnetic
environment - guidance
Not applicable
Portable and mobile RF communications equipment should be used no closer to any part of the Device
,including cables, than the recommended separation distance calculated from the equation appropriate
for the frequency of the transmitter. Recommended separation distances:
d=1.2√P;
d=2√P