Trouble Shooting; Error Messages - Pharmacia Biotech MultiTemp III 18-1106-33 User Manual

Multitemp iii user manual
Table of Contents

Advertisement

5.1 Error
messages
Message on display
Locked display

5. Trouble shooting

Cause
The bath temperature has exceeded
the safety temperature limit.
Too low water level.
The winding of the pump motor is too hot.
The temperature sensor is broken.
The bath liquid has cooled down below
the temperature limit.
The winding of the pump has cooled down Check pump and wait a few minutes until the
below the alarm level.
The temperature sensor is
functioning again.
An undefined fault has occurred.
Undefined cause.
Remedy
Pour in more liquid.
Pour in more liquid.
Check pump and wait a few minutes until the
motor temperature has dropped.
If this fault occurs often, return the unit for
service.
The sensor must be exchanged by service
personnel.
Check the water level in the bath and
adjust if necessary.Restart the unit by
pressing reset.
motor temperature has dropped.
If this fault occurs often, return the unit for
service. Restart the unit by pressing reset.
Restart the unit by pressing reset.
This error may require service personnel.
However, before contacting service personnel,
top up the bath with some water. This can
solve the problem.
Press both arrow keys and the main switch
simultaneously to get
in the display.
Switch off the unit and restart.
Note: The unit is now changed to the
factory set parameters.
5. Trouble shooting
13