Temperature Setting; Temperature Correction Value - Pharmacia Biotech MultiTemp III 18-1106-33 User Manual

Multitemp iii user manual
Table of Contents

Advertisement

4.4 Tempera-
ture setting
4.5 Tempera-
ture
correction
value
2. Use
or
rature. The factory pre-set value is 90.0 °C.
3. Confirm with
al temperature value.
Note: The new value is not saved until
sed.
1. Press
to select the S mode.
2. Use
or
3. Confirm with
returns to the actual temperature value.
Note: The new set value is not saved until
pressed.
The display continuously indicates the actual temperature in
the bath at the position of the sensor. This temperature does
not correspond directly to the temperature in the externally
connected equipment. Under most circumstances this tempe-
rature difference is very small. To adjust the display value to
reflect the temperature in the external system, proceed as fol-
lows:
1. Measure the temperature in the external system.
2. Calculate the temperature difference between this value
and the value displayed.
T
= T
-T
diff
disp
meas
has to be entered.
3. Press
to select the C mode
4. Use
or
- sign. The range is -25.0 to +2.50 °C. The factory pre-set
value is 0.0.
to set the maximum allowed set tempe-
before the display returns to the actu-
to set the working temperature.
before the display automatically
T
is the correction value "c" that
diff
to set the correction value with its + or
4. Operation
has been pres-
has been
11