3. Installation
3.5 Mains
connection
8
liquid pour into a large vessel using hydrostatic pressure. The
bath contains 3 litres of liquid, plus the volume in tubing and
connected equipment. The last millilitres can be removed by
opening the screw on the front of the unit, (see Fig. 1).
Warning! Do not drain hot water. Use protective gloves and
clothing when draining the bath liquid.
Compare the local mains voltage and frequency with the spe-
cifications written on the label on the rear of the unit before
connecting the unit to a grounded mains socket. Deviations
of voltage and frequency must not be more than those speci-
fied in this manual (95 to 132 V/60 Hz, 95 to 125 V/ 50 Hz,
198 to 253 V/50 Hz). Connect the unit to a grounded mains
socket only. The socket must be rated as suitable for the total
power consumption of the unit (1,200 W).