Wissner-Bosserhoff Eleganza 1 User Manual And Technical Description

Wissner-Bosserhoff Eleganza 1 User Manual And Technical Description

Hospital bed for standard care
Table of Contents

Advertisement

Quick Links

User manual
Eleganza 1
Hospital bed for standard care
D9U001GT0-0201
Version: 01
Print date: 2017-06
and technical description

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Eleganza 1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Wissner-Bosserhoff Eleganza 1

  • Page 1 User manual and technical description Eleganza 1 Hospital bed for standard care D9U001GT0-0201 Version: 01 Print date: 2017-06...
  • Page 2 It is for this reason that the contents of this manual may indicate differences from the current manufacture of the product. 2/68 Product: Eleganza 1 Version: 01...
  • Page 3: Table Of Contents

    8.5 Eleganza 1 (1GTA) – Split Plastic Side Rails ...................22 8.6 Eleganza 1 (1GTD) – Wooden Bed Ends ..................23 8.7 Eleganza 1 (1GTP) – Single collapsible side rails controlled from top ..........24 8.8 Eleganza 1 (1GTP) – Single collapsible side rails controlled from bottom ........25 8.9 Bed Ends ............................26...
  • Page 4 17 EC Declaration of Conformity ..................60 18 Technical Specifications ....................61 18.1 Mechanical Specification ......................61 18.2 Electrical Specification .........................62 18.3 PB 4X Electronic System of Eleganza 1..................63 19 Log 19.1 Delivery Log ..........................67 19.2 Service and Maintenance Log .....................68 4/68...
  • Page 5: Symbols

    Perform this step. Result, if necessary. 1.2.2 Lists Structure of bulleted lists:  List level 1  List level 2 Structure of numbered lists: List level 1 List level 1 List level 2 List level 2 5/68 Product: Eleganza 1 Version: 01...
  • Page 6: Symbols And Labels On The Product

    Jack for attachment of conductor for potential equalisation Safe Working Load Warning against crushing or trapping Maximum weight of patient Use mattress recommended by manufacturer Weight of bed Do not put any objects here Designation of hospital bed for adults 6/68 Product: Eleganza 1 Version: 01...
  • Page 7: Serial Label With Udi

    Serial label with UDI 7/68 Product: Eleganza 1 Version: 01...
  • Page 8: Safety And Dangers

    2 Safety and Dangers SIGNAL WORDS! Eleganza 1 bed should be left in its lowest position when the patient is unattended in order to reduce risk of injury due to falls! SIGNAL WORDS! Siderails of Eleganza 1 should be located in the „up“ position to reduce the risk of the patient...
  • Page 9: Additional Safety Instructions For Psychiatric Bed

    Periodically check if the bed placed on the psychiatric department is not damaged. In case of any damage of the bed, replace the damaged parts.  Always check grounding of the bed in case the bed is equipped with grounding chain. 9/68 Product: Eleganza 1 Version: 01...
  • Page 10: Conditions Of Use

    The bed is designed for use in rooms for medical purposes. Electrical installations must therefore meet local standards laying down the necessary conditions for electrical installations.  Disconnect the bed from the mains in exceptional cases (i.e. a storm). 10/68 Product: Eleganza 1 Version: 01...
  • Page 11: Standards And Regulations

    EN ISO 10993-10:2013  93/42/EEC  2011/65/EU The manufacturer adheres to a certified quality management system in compliance with the following standards:  EN ISO 9001:2015  EN ISO 14001:2015  EN ISO 13485:2012 11/68 Product: Eleganza 1 Version: 01...
  • Page 12: Functioning

    Always conduct the risk assessments required for the selection of suitable equipment. Eleganza 1 is a standard care hospital bed. It is designed for supporting the patient in several positions such as horizontal, Trendelenburg etc. The bed is not designed for intensive care. The stable undercarriage with 4 castors makes it possible for the bed to be controlled by one person.
  • Page 13: Scope Of Delivery And Bed Variants

    Notify the carrier and supplier of any deficiencies or damages immediately as well as in writing or make a note on the delivery note. Bed Variants Eleganza 1 – Model 1GT Hospital Bed 5.2.1 Features - Eleganza 1 Model 1GT Hospital Bed (model no. see product label): s = standard o = optional  Mattress Platform ...
  • Page 14 Eleganza 1 – Model 1GD Nursing Bed 5.2.2 Features - Eleganza 1 Model 1GTD Nursing Bed (model no. see product label): s = standard o = optional  Mattress Platform  mattress platform consisting of removable plastic parts (s) ...
  • Page 15: Applied Parts Type B

    Eleganza 1 – Model 1GP Psychiatric Bed 5.2.3 Features - Eleganza 1 Model 1GTP Psychiatric Bed (model no. see product label): s = standard o = optional  Mattress Platform  mattress platform consisting of fixed HPL parts (s) ...
  • Page 16: Setup

    There is no mains switch on the bed, i.e. the mains cable is the only means to isolate the bed from the mains.  Ensure the mains cable is always accessible.  The plug on the mains cable should only be changed and maintained by qualified and trained service technicians authorised by the manufacturer. 16/68 Product: Eleganza 1 Version: 01...
  • Page 17: Battery Activation

    Isolating Foil Check if isolating foil is complete and undamaged as shown: If isolating foil is damaged, contact the manufacturer’s service department immediately. NOTE: It is recommended to wear gloves when removing the isolating foil. 17/68 Product: Eleganza 1 Version: 01...
  • Page 18: Assembly

    Make sure the connectors match.  Make sure there is no possibility of accidental disconnection. Before moving the bed:  Disconnect the patient from the intravascular or intracardiac device.  Disconnect the potential equalisation connector. 18/68 Product: Eleganza 1 Version: 01...
  • Page 19: Lockable Box For Mains Cable (Psychiatric Version)

    Enclose the magnetic key to the lock and gently push.  Grab magnetic key and lock.  Remove the lock by pulling. Fig. 3 Control elements of the box Fig. 4 Unlocking the box 19/68 Product: Eleganza 1 Version: 01...
  • Page 20: Mattress Platform

    Ensure that mattress platform sections are fitted correctly by trying pull the section up with little manpower. Fig. 6 Label on mattress platform parts (H = head, F = foot, R = left, L = right) 20/68 Product: Eleganza 1 Version: 01...
  • Page 21: Eleganza 1 (1Gta) - Single Collapsible Side Rails

    Eleganza 1 (1GTA) – Single Collapsible Side Rails Fig. 7 Overview Eleganza 1 – Single collapsible side rails 1. Foot end 2. Single collapsible side rails 3. Side rail release mechanism 4. Head end 5. Handset 6. Castors 7. Central castor system lever 8.
  • Page 22: Eleganza 1 (1Gta) - Split Plastic Side Rails

    Eleganza 1 (1GTA) – Split Plastic Side Rails Fig. 8 Overview Eleganza 1 – Split plastic side rails 1. Foot end 2. Split collapsible side rails 3. Control element, integrated in side rail 4. Head end 5. Castors 6. Side rail release mechanism 7.
  • Page 23: Eleganza 1 (1Gtd) - Wooden Bed Ends

    Eleganza 1 (1GTD) – Wooden Bed Ends Fig. 9 Overview Eleganza 1 - Wooden Bed Ends 1. Foot end 2. Side rail release mechanism 3. Mattress base 4. Head end 5. Castors 6. Control element, integrated in side rail 7. Single collapsible side rail 8.
  • Page 24: Eleganza 1 (1Gtp) - Single Collapsible Side Rails Controlled From Top

    Eleganza 1 (1GTP) – Single collapsible side rails controlled from top Fig. 10 Eleganza 1 (1GTP) – Single collapsible side rails controlled from top 1. Foot end 2. Side rail release mechanism 3. Single collapsible side rails controlled from the top 4.
  • Page 25: Eleganza 1 (1Gtp) - Single Collapsible Side Rails Controlled From Bottom

    Eleganza 1 (1GTP) – Single collapsible side rails controlled from bottom Fig. 11 Eleganza 1 (1GTP) – Single collapsible side rails controlled from bottom 1. Foot end 2. Mattress platform consisting of fixed HPL parts 3. Single collapsible side rails controlled from the bottom 4.
  • Page 26: Bed Ends

    Material damage due to excess load!  Ensure nobody sits on the bed ends. NOTE: The bed ends may be delivered in several decor colour variants. Fig. 12 Locking the bed end 1. Locked 2. Unlocked 26/68 Product: Eleganza 1 Version: 01...
  • Page 27 Lock safety levers on corner posts. Remove bed ends as follows:  Unlock safety levers on corner posts.  Pull bed end upward. Fig. 13 Installing the bed end 1. Right (coloured panel outside) 2. Wrong (coloured panel inside) 27/68 Product: Eleganza 1 Version: 01...
  • Page 28: Operation

    The manufacturer provides a 6-month warranty for the full function of batteries.  Check the battery functionality at least once a month in accordance with the user and service manuals and have the battery changed if necessary. 28/68 Product: Eleganza 1 Version: 01...
  • Page 29 In case the battery cover or control section is deformated by heat  Unplug the bed from the mains.  Do not use the bed (see. Removing the Bed from Service.  Contact service of the manufacturer. 29/68 Product: Eleganza 1 Version: 01...
  • Page 30: Status Faulty Battery

    Press the Thigh Rest Up + Thigh Rest Down + Trendelenburg Position buttons at the same time and hold them for three seconds.  The battery is deactivated. NOTE: Try some functions to ensure the battery is deactivated. 30/68 Product: Eleganza 1 Version: 01...
  • Page 31: Manipulation

    Disabling individual functions on the supervisor control panel affects all of the control elements. If the bed does not react to individual position settings:  Check whether the function is disabled on the supervisor control panel. 31/68 Product: Eleganza 1 Version: 01...
  • Page 32: Miniacp (Supervisor Board)

    10. Reverse Trendelenburg Button 11. CPR Button 12. Trendelenburg Button (optional) To set positions:  Activate the keypad by pressing the GO button.  Press and hold corresponding button until required position is reached. 32/68 Product: Eleganza 1 Version: 01...
  • Page 33: Acp - Supervisor Control Panel

    15. Mains Power LED 16. Central STOP Button 17. Activating GO Button To set positions:  Activate the keypad by pressing the GO button.  Press and hold corresponding button until required position is reached. 33/68 Product: Eleganza 1 Version: 01...
  • Page 34 The respective LED flashes 1 to indicate the lock. NOTE: The individual functions are locked in the central control panel, the satellite control, the handset, and the side rail control. The foot controllers are locked separately. 34/68 Product: Eleganza 1 Version: 01...
  • Page 35 The CPR position is for resuscitating the patient in an emergency. Settings after pressing and holding the CPR button (11):  Mattress platform moves into a horizontal position. NOTE: For quick mechanical positioning, see CPR Backrest Release. 35/68 Product: Eleganza 1 Version: 01...
  • Page 36: Handset

    If required, prevent the patient from adjusting the bed as follows:  Disable functions.  Unplug handset (if the bed is equipped with Plug and Play).. ® NOTE: The handset can be plugged into another LINET hospital bed with the Plug and Play slot. 36/68 Product: Eleganza 1 Version: 01...
  • Page 37: Satellite Control Panel

    If required, prevent the patient from adjusting the bed as follows:  Move the satellite panel out of the patient’s reach. -or-  Disable functions. NOTE: The satellite control panel can be fixed on the right or left side of the bed. 37/68 Product: Eleganza 1 Version: 01...
  • Page 38: Control Element Integrated In Side Rail

    10.5 Control Element Integrated in Side rail It is possible to equip the Eleganza 1 bed with integrated illuminated controllers in the side rails. The optional control element is attached to the side rail. The controllers are positioned on both sides of the bed.
  • Page 39: Foot Control

    3. Protection Frame against Unwanted Activation 4. Foot Switch Raise Mattress Platform Set the position as follows:  Press foot switch for activation (mini GO function).  Press and hold foot switch until desired position is reached. 39/68 Product: Eleganza 1 Version: 01...
  • Page 40: Cpr Backrest Release

    Set the CPR position as follows:  Press and hold lever upwards.  Press the backrest down using:  Mattress guard handle.  Upper part of the side rails (in case the split side rails are in upper position) 40/68 Product: Eleganza 1 Version: 01...
  • Page 41: Side Rails

    To fold side rail up push the side rail up. Side rail will click into place and lock automatically (you will hear audible ‘click’). Fig. 22 Releasing mechanism of the single collapsible side rails with bottom control 41/68 Product: Eleganza 1 Version: 01...
  • Page 42 Repeat on the other side of the side rail.  To fold side rail up push the lift the side rail up. Side rail will click into place and lock automatically. Fig. 24 Releasing mechanism of the full length side rails 42/68 Product: Eleganza 1 Version: 01...
  • Page 43: Castor Control And Bed Transport

    Brake all castors of the bed by pressing the lever 1 by your foot – castor is locked.  Unbrake all castors of the bed by lifting up the lever 2 by your foot – castors are unlocked. 43/68 Product: Eleganza 1 Version: 01...
  • Page 44: Accessories

    NOTE: The manufacturer assumes no responsibility for the use of accessories not approved by the manufacturer. NOTE: All accessories conform to IEC 60601-2-52. The following accessories are suitable for the Eleganza 1:  Lifting pole  powder-coated  chrome-plated ...
  • Page 45 Use exclusively infusion stands with 4 hooks for hanging IV bags or baskets for intravenous solutions.  Ensure that the weight-bearing capacity of the 4 hooks is not exceeded. Capacity per hook: 5 kg  Ensure that the infusion stand’s weight-bearing capacity is not exceeded. 45/68 Product: Eleganza 1 Version: 01...
  • Page 46 Mattress Eleganza 1 bed is designed for the following mattress dimensions:  200 cm x 90 cm x 14 cm The manufacturer recommends the use of the following mattress systems on the Eleganza 1 bed:  Efecta - 4PRB7GA7300 ...
  • Page 47 Dismount Eleganza Protector® as follows:  Press and hold the releasing button (1).  Push the protector up. ® NOTE: Protector cannot be used if the bed is equipped with plastic E3 bed ends. 47/68 Product: Eleganza 1 Version: 01...
  • Page 48 Secure the screws with the plastic rosette nuts and tighten (Fig. 32). The rosette nuts are pointed outside from the bed. ®  Check if the Extender is secured correctly by pushing it sideways. 48/68 Product: Eleganza 1 Version: 01...
  • Page 49  Secure the screws with the plastic rosette nuts and tighten (Fig. 34). The rosette nuts are pointed outside from the bed.  Check if the Extender® is secured correctly by pushing it sideways. 49/68 Product: Eleganza 1 Version: 01...
  • Page 50: Cleaning And Disinfection

    Electronic parts that may be contaminated clean carefully and only their outer side. Never open those connectors due to cleaning or disinfection. Those components should not be exposed to prolonged or continuous exposure to moisture.  Let the bed dry completely after cleaning or disinfection process. 50/68 Product: Eleganza 1 Version: 01...
  • Page 51  After drying the bed place the mattress back on the mattress platform.  After drying the bed check functions of the bed. 51/68 Product: Eleganza 1 Version: 01...
  • Page 52: General Instructions For Cleaning And Disinfection

    Mattress should be periodically checked for cracks, holes or cracks that may affect the integrity, water resistance or resistance to infections of the cover. Contact the service department of the manufacturer according to scope of damage to cover. 52/68 Product: Eleganza 1 Version: 01...
  • Page 53: Choosing Of Detergents Or Disinfections

    Risk of damaging the bed due to use of incorrect detergent!  Always consult choosing of the detergent and its dilution with manufacturer of the detergents according to the material table below. Changing Part of bed – Eleganza 1 Material (*) Daily C&D Complete C&D patient C&D Side rails (stickers) ...
  • Page 54: Troubleshooting

    Position head or foot board correctly. Defective mechanism Notify service department. 13.1 Sound signals of the bed Battery is almost discharged – the signal sounds after each press of the handset. 54/68 Product: Eleganza 1 Version: 01...
  • Page 55: Maintenance

    Information about spare parts is available from:  Customer service  Sales  Our technical support department 14.1.2 Completeness  Perform a visual check (with delivery note if necessary).  Have any missing parts replaced. 55/68 Product: Eleganza 1 Version: 01...
  • Page 56: Functioning

    Have the brakes adjusted if necessary.  Have any defective castors replaced. 14.2.3 Accessories  Check all accessories (for example, lifting pole, side rails, infusion stand, etc.) are working properly.  Replace if necessary. 56/68 Product: Eleganza 1 Version: 01...
  • Page 57 In accordance with the Medical Devices Operator Ordinance, the operator is required to perform a technical safety check on the hospital bed every 12 months. The procedure for performing the safety check is stipulated in EN 62353. 57/68 Product: Eleganza 1 Version: 01...
  • Page 58: Disposal

    Dispose of the bed or its components in accordance with local laws and regulations:  After using the bed  Following maintenance and installation work  Hire an approved waste disposal company for disposal. 58/68 Product: Eleganza 1 Version: 01...
  • Page 59: Warranty

    Failures and errors caused by incorrect use and external effects are not covered. Justified complaints will be fixed free of charge during the warranty period. Proof of purchase, with the date of purchase, is required for all warranty service. Our standard terms and conditions apply. 59/68 Product: Eleganza 1 Version: 01...
  • Page 60: Ec Declaration Of Conformity

    17 EC Declaration of Conformity 60/68 Product: Eleganza 1 Version: 01...
  • Page 61: Technical Specifications

    250 Kg  without battery 200 Kg Max. mattress height split plastic side rails 15.9 cm single collapsible side rails 16.7 cm side rails equipped with Extender 26.7 cm psychiatric bed 15 cm 61/68 Product: Eleganza 1 Version: 01...
  • Page 62: Electrical Specification

    (custom voltage, different mains plugs). Identification of applied parts (Type B):  supervisor panel  handset  side rails  bed ends  mattress platform 62/68 Product: Eleganza 1 Version: 01...
  • Page 63: Pb 4X Electronic System Of Eleganza 1

    Eleganza 1 requires special preliminary measures pertaining to EMC that necessitates installation and commissioning in conformity with the EMC information given in this Manual. Eleganza 1 is intended for the application in electromagnetic environment as specified below. The customer or user of the bed is responsible for these requirements are met.
  • Page 64 (50/60 Hz) frequency conform to IEC 61000-4-8 the normal levels of commercial or hospital environments. NOTE U refers to the AC mains voltage before the test level is applied. 64/68 Product: Eleganza 1 Version: 01...
  • Page 65 The field intensity in the entire frequency range from 150 kHz to 80 MHz should be lower than 3 V/m. NOTE With 80 MHz and 800 MHz, the higher frequency range is applicable. NOTE The absorption and reflection of buildings, objects and people will influence electromagnetic propagation 65/68 Product: Eleganza 1 Version: 01...
  • Page 66 Eleganza 1 Medical Bed The Eleganza 1 Medical Bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Eleganza 1 Medical Bed can help prevent...
  • Page 67: Log

    19.1 Delivery Log Order number: Customer: Model number: Serial number: Delivery date: Delivered by: I hereby confirm personnel have been trained to correctly operate the bed. Date: Customer‘s Signature and Seal: Supplier’s Signature and Seal: 67/68 Product: Eleganza 1 Version: 01...
  • Page 68: Service And Maintenance Log

    19.2 Service and Maintenance Log Description of Service Date Performed By 68/68 Product: Eleganza 1 Version: 01...