Wissner-Bosserhoff Carisma Bed Instruction & Safety Manual Please read this manual before operation, for maintenance instructions and safe usage The enabling power of applied knowledge 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
Page 2
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual Index Table of Contents ..........................2 Important Safety Note ........................3 Notes on Instructions for use / Explanation of Symbols ...............4 Area of Use / Intended Use ......................4 Setting up the Bed ........................5 Technical Data for the Bed / Nameplate ..................5 5.1.
Page 3
Setting up the Nursing Home Bed (Safety Clearances) ................50 Cleaning and Disinfection of the Bed ....................50 Appendixes: Appendix A: Spare parts list carisma 09/10 Appendix B: Nameplate – explanation of details using an example Appendix C: Important safety and usage notes for nursing home beds...
Any other applications must be agreed beforehand in writing with the company Wissner-Bosserhoff GmbH. The product must be used as a utility for nursing, and is subject to the regulations of the relevant professional organisation.
fl oors, porous and soft stone fl oors, carpets with foam backing, soft linoleum or similar fl oor covering. If in doubt, please contact your fl ooring company and the Technical Customer Service of Wissner-Bosserhoff GmbH. •...
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 5.1. Electromotor Patient Surface • Electrical connection: 230 V 1,25 A 50 Hz • Protection type IPX4 • Class of protection • Output voltage 24 V / 32 V depending on model variant (please read nameplate) •...
6. Product Safety / Extract From Applied Standards The product bears the mark and thus complies with the German and European safety regulations to be applied to the product. Standard Comment Standard Comment Medical product acc. to MDD CE- Symbol DIN 33402 Body mass of people 93/42/EEC...
Caution! All electrical functions could have faults - unless you lock them and just release them for the duration of the function. Caution! Other devices that emit electromagnetic radiation (e.g. radio telephones) can have an infl uence on electrical bed functions. For this reason in general it applies that bed functions not required have to be locked for reasons of safety.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 8.2. Legs Up Position (Lu) The adjustment is made via the above mentioned buttons of the handset. Caution! For reasons of safety, this position may only be released by medical staff, or only “under medical supervision“! The handset keys are released via the safety key!
Caution! Because two power units independent of one another are used for the height adjustment in the “carisma“ model, it is important, depending on the adjustment cycle, to move the patient surface at least once per day to the end position (highest or lowest position).
One of these safety precautions to guarantee initial fault tolerance is the control box. When using initial fault tolerance handsets, such as the Carisma 07 NPF with integrated locking mode, the control box is not necessary. Handling the locking device with the handset is described in the Handset chapters.
Caution! The handset should be in the so-called “Resident Mode“. This means that the comfort sitting position (CS) and legs up position (LU) are unavailable! To release the above-mentioned positions, rotate the locking switch on the rear of the handset. (Chapter: Handset –...
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 13. Transport / Moving Bed Bring bed into mobile position and move bed (see Chapter Brakes / Moving). Hold only the head and foot ends when moving the bed (see Fig. 2). Under no circumstances may the bed be pulled/pushed...
These are explained in more detail in the following chapter. 14.1. Handset “Carisma 07 Jpf“ Function Description Caution: set the locking switch before you start. (13.4)
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 14.2. Handset – Rear Side With Locking Function Locking switch for various modes: By turning the safety key, 3 modes can be adjusted. Locking mode = All functions are locked Care mode =...
Initial fault tolerance of the handset: When operating the keys, a short signal can be heard. This means that the handset function is being performed correctly. If a continuous sound is heard when the key is not pressed, either the handset function is locked by the control box (optional) or there is an error.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 14.4. Braking/ Moving Caution! : Due to the intended use of the bed, its movability is limited. The bed is not intended to transport patients, but only to move it to a fl oor suitable for it. Normally a level fl...
Page 19
14.6. Adjusting Side Rails (Release Catch 09) Side rails up/down Operation: Lowering (A): Pull up side rail (C) and simultaneously press in button (D) and hold. Slowly let side rail glide into the lowest position (A). Raising (B): Hold side rail (C) on the lower edge and pull upwards (towards B) until the button (D) has clicked into position.
Page 20
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 14.7. Adjusting Side Rails (Model 09) Raising/Lowering side rails Operation: 1st level: Side rail is in lowered position. 2nd level: To restrict the patient surface, simply pull the handrail of the side rail upwards until a “click“...
Caution! When lowering /raising the side sections (side rails) you must be very careful not to get fi ngers, hands or other body parts trapped between the side rails and the patient surface frame. Caution! Restless or disoriented patients/residents could become entangled between the safety gaps of the side rails.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 15.2. Infusion Holder (Accessories) The infusion holder (like the lifting pole) is fi xed to the socket provided on the patient surface frame. The infusion stand is only intended for attaching infusions, not for attaching infusion pumps or similar.
15.4. Mattress (Accessories) Our Universal mattresses (dimensions: 87x 200x12cm) have special incisions on their lower side. When the mattress is laid on the bed, it must be ensured that the incisions match the rotation points of the patient surface. Only in this way can an optimum lying contour be guaranteed.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 15.6. Handset Holder (Accessory) The handset holder (B) is used to place the handset where the resident can reach it. It is placed in the accessories holder (A) Fig.23. Positioning options Caution! Since the handset is fl exible, it may not be used as assistance in getting up or as a handle.
Page 25
15.7. Side Rail Extension For 06/09 Side Rail (Accessories) 1. Side rail 2. Tilting arm 3. Mushroom head bolt 4. Clamping lever Assembly: 1. Pull up the side rails of the nursing home bed to the highest position. 2. Release the clamping lever (Fig. 27, Item 4) by turning it 90° and twisting out the mushroom head bolt 3.
Page 26
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual Caution! For side rail protections with extensions, please use the available side protection cushioning (accessory), which is an important additional safety and protection feature. 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
16. Cleaning And Disinfection 16.1. Initial Cleaning Prior to fi rst-time use the nursing home bed must be cleaned in accordance with the IP protection level. Identifi cation Protection level for protection protection level for protection Cleaning and letter from contact and impurities from water Disinfection Designation of extent of...
We have used materials of the highest quality to create a product for you with a high level of benefi ts and homely real wood character. In order that you can gain pleasure for as long as possible from your Wissner-Bosserhoff product, we recommend that you clean with a slightly damp, soft cloth.
For particularly stubborn stains, commercially available, environmentally friendly all-round household cleaners can of course be used. The furniture surfaces of Wissner-Bosserhoff products meet the requirements of Standard DIN 68861 - 1A for furniture surfaces subjected to chemicals, and thus have an excellent resistance to normal pollution.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 17. Help In Case Of Malfunctions Malfunction Possible Cause Remedy E- Motor adjustments not • Mechanics are locked Check moving parts and made correctly remove any foreign bodies E- Motor adjustments are • Handset is defective...
Page 31
The bed may only be operated by persons who can guarantee correct handling because of their training or knowledge and practical experience. Before operating the bed, the person doing this must be satisfi ed with the functional safety and the proper condition of the bed. The bed must be subject to a function test, possibly daily or with each change of shift, so that it is guaranteed that the bed can be used as intended without danger for the patient or the operator.
Page 32
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual Since the system can be run via battery operation, it is not enough just to pull out the plug in case of the motors functioning incorrectly. If an incorrect function occurs, please switch off functions via the control box and contact our Customer Service (battery operation option).
Page 33
Cleaning of the bed may only be carried out with a damp cloth. If the bed is not used properly, dangers can occur through: • Unauthorised operation of electrical functions during the uncontrolled positioning of the bed, Using the bed for children under the age of 12, •...
Page 34
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual • Cleaning the bed may only be carried out in accordance with the specifi cations in Section “Cleaning and Disinfection“ of these instructions for use, since increased risks could otherwise occur for which the manufacturer can take no responsibility.
Page 35
• Furthermore, electrically-operated hospital and nursing home beds are electrical resources, for the safety of which the employer is responsible. The monitoring function for this obligation is the responsibility of the German professional association for health service and welfare care (BGW) and the German industrial inspectorate.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual 20. Maintenance And Servicing The nursing home beds of Wissner-Bosserhoff GmbH are generally designed for a useful life of 10-15 years. The beds here are extremely low in maintenance. This is because attention was paid during the development of the products to ensure that maintenance work was reduced to a minimum, with low operating costs.
Test of functionality of castors (moving – braking) and the removal of possibly accumulated dust, threads, hair or similar that may have been caught in the castors. If necessary the castor brakes must be readjusted. Please ask Wissner-Bosserhoff Customer Service. 21. Useful Life If used, cleaned, serviced and repaired correctly, a useful life for our products of 10 to 15 years can be expected.
Spare part lists, current price lists and servicing instructions with exploded diagrams can be requested if required from Wissner-Bosserhoff, Technical Customer Service, by specifying the nameplate data of the bed or the relevant article number, order number and delivery date.
(see Appendix B of these instructions for use) so that the Wissner-Bosserhoff customer consultant can be notifi ed of these. If these data are not available, guarantee services in particular cannot be treated free of charge.
Page 41
Appendix A: Spare Parts List “Carisma“ 09/10 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
Page 42
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual Enter important ordering details from the bed nameplate! Article No. Series No. Enter important ordering details from the control motor name plate! Article No. Series No. 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...
The conditions of the respective instructions for use of the product shall apply: The following points must be observed: The nursing home beds of Wissner-Bosserhoff GmbH are intended exclusively for the occupancy of persons from the age of 12 and above. This must be ensured, since persons with a body size or weight deviating from the normal average are exposed to an increased risk under certain circumstances.
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual The operation of nursing home dates is allowed exclusively from mains sockets installed in the building that are up-to-date and that are inspected regularly for the correct function of safety installations (RCD protection, insulation error detectors, power fuse, etc.) in accordance with valid regulations.
Page 47
The safety notes in the specifi c product instructions for use must be observed, since hazards could otherwise occur. When using the side guards, particular care and caution should be taken: - side guards are used fi rst and foremost to protect the patient from rolling out. On no account may the side rails be used as a means to hold the patient fi...
Page 48
fi gure of eight loop and held together with a strong rubber band. Alternatively the Wissner-Bosserhoff GmbH mains cable hook can be used. (see following diagrams). Under no account may fi xed cable binders be used, since these could damage the mains cable if it is subject to unintentional tension.
USE OF ORIGINAL ACCESSORIES AND RELEASED PRODUCT COMBINATIONS The use of accessories and combinations of products not authorised for the product can lead to serious hazards. This is why only accessories released by Wissner-Bosserhoff GmbH and products released for combination may be used. Examples: Using lifting poles with an insuffi...
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual Setting Up The Nursing Home Bed (Safety Clearances) The bed must be set up in such a way that when implementing all adjustment functions of the bed, collision of bed parts with other objects in the room or with parts of the room is not possible (e.g.
Page 51
Appendix D: Battery Operation (Option) – Additional Notes As a special fi tting, the bed power unit can be equipped with an additional battery function allowing the bed to be moved multiple times with its electrical functions independently of the mains supply. The number of adjustments possible depends on the charge status of the battery.
Page 52
Wissner-Bosserhoff Carisma Bed, Instruction & Safety Manual PO Box 82 Bentley WA 6982 Authorised Distributor T 08 9351 7500 F 08 9350 5299 info@unicarehealth.com.au www.unicarehealth.com.au 08 9351 7500 info@unicarehealth.com.au www.unicarehealth.com.au...