Examination; Important Information - Braster Pro User Manual

Breast examination system
Hide thumbs Also See for Pro:
Table of Contents

Advertisement

7. EXAMINATION

7.1. IMPORTANT INFORMATION

Do not use the device while it is being charged.
REMEMBER
Do not perform an examination in a room where the
temperature is lower than 20˚C or higher than 25˚C.
Due to changes in breast physiology during the menstrual
cycle, if you wish to compare the subsequent results of the
same patient, select the examinations performed on a similar
day of the menstrual cycle.
Prior to the examination, please make sure that the
patient:
1. On the day before the examination:
• avoided procedures which may have increased blood
circulation in the breast area (physiotherapy, massages, etc.).
2. On the day of the examination:
is 18 years of age or older;
avoided procedures which may have increased blood
circulation in the breast area (physiotherapy, massages, etc.);
did not take a bath or a shower within approximately two
hours before the examination;
did not use peelings on her breasts and/or neckline and did
not perform armpit depilation, as this may have irritated her
skin;
refrained from sources of heat or cold, such as heaters or air
conditioners, for at least thirty minutes before and during the
examination;
avoided intense physical effort for approximately two hours
prior to the examination;
did not use cosmetics and did not wet the skin on her breasts
within two hours prior to the examination – her skin should
23

Advertisement

Table of Contents
loading

Table of Contents