Problem
Symptom
E6 shows
E10 or E11 shows
E20 shows
Error message
E21 shows
EExx shows on the display
13
specifications
Power supply
Display mode
Measurement mode
Measurement range
English
44
Check this
The voltage of the batteries is
too low during the inflation
The monitor detected motion
while measuring.
The measurement process
does not detect the pulse
signal.
Measure incorrectly.
A calibration error occurred.
Battery powered mode: 2 x AAA batteries (3V DC)
Digital LCD V.A. 32 mm x 45 mm
Oscillographic testing mode
Rated cuff pressure: 0 mmHg~300 mmHg (0 kPa ~40 kPa)
Measurement pressure: SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40 mmHg~130 mmHg (5.3 kPa~17.3 kPa)
Pulse value: (40-199) beat/minute
Remedy
Measure again. If E6 appears again, please replace with
new batteries.
Movement can affect the measurement. Relax for a
moment and then measure again.
Loosen the clothing on the wrist and then measure
again.
Relax for a moment and then measure again.
Retake the measurement. If the problem persists,
contact the retailer or our customer service for
further assistance. Refer to the warranty for contact
information and return instructions.