9.1.2
Flow Cells
Type
Standard flow cell,
SST
Semi-analytical
flow cell, SST
Semi-micro flow
cell, SST
Standard bio-
compatible flow
cell, PEEK
Semi-micro bio-
compatible flow
cell, MP35N
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
The flow cell performances are as follows:
Path length
Illuminated
volume
10 mm
13 µL
7 mm
5 µL
7 mm
2.5 µL
10 mm
13 μL
7 mm
2.5 µL
Pressure
Max. eluate
limit
temperature
limit
12 MPa
80 °C
12 MPa
80 °C
12 MPa
80 °C
5 MPa
60 °C
12 MPa
80 °C
9
•
Specifications
Materials
Bio-
in the flow
compatibility
path
Stainless
No
steel, fused
silica, PTFE,
PEEK,
titanium
Stainless
No
steel, fused
silica, PTFE,
PEEK,
titanium
Stainless
No
steel, fused
silica, PTFE,
PEEK,
titanium
Fused
Yes
silica, PEEK
MP35N,
Yes
titanium,
fused silica,
PTFE, PEEK