The user may perform manual calibration of the meter, provided they have standard
resistors, minimum grade 0.05%, of the following nominal values: 0,2mΩ, 2mΩ, 20mΩ,
200mΩ, 2Ω, 20Ω, 200Ω and 2kΩ. Calibration is performed separately for each range.